SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION recall records

6 reported ongoing

6 records in this archive. Latest source date Oct 16, 2025.

FDA enforcement status can lag behind the recall lifecycle. An empty search or a reported termination is not a personal product-safety determination.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

Records and reported status

FDA device enforcement

Ceiling Type X-Ray Tube Support Models CH-200 and CH-200M used with RADspeed, SonialVision and FLUOROspeed Systems The ceiling type x-ray tube support CH-200 is used to hold the...

Due to an issue (potential breakage) with the shaft for the mounting of the X-ray tube assembly that could potentially result assembly coming off the rotating shaft.

Class IIReported ongoing

Distribution: Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, N...

Official source for Ceiling Type X-Ray Tube Support Models CH-200 and CH-200M used with RADspeed, SonialVision and FL...
FDA device enforcement

Trinias unity, Digital Angiography System with Catheterization Table, Model: KS-100 This system consists of the following main components: C arm, catherization table, image proc...

If digital angiography system is used with a specific catheterization table and the operator directs simultaneous movement of the table and the C-arm using direct memory, then direct memory button is released, the table may continue to move, which the opera...

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of MA, IL, SC, KY, SC, OH, TN.

Official source for Trinias unity, Digital Angiography System with Catheterization Table, Model: KS-100 This system c...
FDA device enforcement

SONIALVISION G4 is a multi-purpose X-ray R/F system. The equipment is suitable for radiographic and fluoroscopic examinations, including general radiography and pediatric examin...

X-ray R/F system software with certain versions of the control board may occasionally fail to display images when switching from Fluoroscopy mode to Radiography mode, which may lead to the possibility of delay or cancellation of the examination.

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of CA, CT, MA, WV, OH, NY, IA, NE, MN, IL, FL, NC, HI, LA, MD, MT, WA, AR, TX, TN, CO.

Official source for SONIALVISION G4 is a multi-purpose X-ray R/F system. The equipment is suitable for radiographic a...
FDA device enforcement

The FLUOROspeed is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a flat detector, it can perform a range of appli...

X-ray R/F system software with certain versions of the control board may occasionally fail to display images when switching from Fluoroscopy mode to Radiography mode, which may lead to the possibility of delay or cancellation of the examination.

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of CA, CT, MA, WV, OH, NY, IA, NE, MN, IL, FL, NC, HI, LA, MD, MT, WA, AR, TX, TN, CO.

Official source for The FLUOROspeed is intended to be used as a universal diagnostic imaging system for radiographic ...
FDA device enforcement

Over-Head Ceiling Tube Support, CH-200, CH-200M, Used with RadSpeed and RadSpeed PRO, Sonial Vision safire, Versa and G4 units, FluoroSpeed 300 and X1 units

Eight fixing bolts at the base of the X-ray ceiling tube column assembly may be loose or missing so service is required to install a cable, bracket, and label behind the bracket, indicating when service is required, to eliminate the risk of the device falling.

Class IIReported ongoing

Distribution: U.S. Nationwide distribution.

Official source for Over-Head Ceiling Tube Support, CH-200, CH-200M, Used with RadSpeed and RadSpeed PRO, Sonial Visi...

Reasons in these listed records

  • Due to an issue (potential breakage) with the shaft for the mounting of the X-ray tube assembly that could potentially result assembly coming off the rotating shaft.
  • If digital angiography system is used with a specific catheterization table and the operator directs simultaneous movement of the table and the C-arm using direct memory, then direct memory button is released, the table may continue to move, which the operator can stop by pres...
  • X-ray R/F system software with certain versions of the control board may occasionally fail to display images when switching from Fluoroscopy mode to Radiography mode, which may lead to the possibility of delay or cancellation of the examination.
  • Due to capacitor manufacturing issue, their is a potential that Flat Panel Device may have abnormal display or no function.

These examples come from this result page and are not a ranking of every recall reason.

Checking SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION

Is there a recall on shimadzu corporation medical systems division right now?

This archive contains 6 records, including 6 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking