Smiths Medical ASD Inc. recall records

138 reported ongoing

138 records in this archive. Latest source date Apr 10, 2025.

FDA enforcement status can lag behind the recall lifecycle. An empty search or a reported termination is not a personal product-safety determination.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

Records and reported status

FDA device enforcement

Medfusion Syringe Pump, Model 3500-500, v3, v4, v5, and v6

A force sensor in the occlusion detector may drift out of calibration leading to increased occlusion detection times, false occlusion alarms, or a System Failure Alarm. If the force sensor calibration shift is large enough, the pump will display a System Fa...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the country of Canada.

Official source for Medfusion Syringe Pump, Model 3500-500, v3, v4, v5, and v6
FDA device enforcement

Medfusion Syringe Pump, Model 3500VX-500, v3, v4, v5, and v6

A force sensor in the occlusion detector may drift out of calibration leading to increased occlusion detection times, false occlusion alarms, or a System Failure Alarm. If the force sensor calibration shift is large enough, the pump will display a System Fa...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the country of Canada.

Official source for Medfusion Syringe Pump, Model 3500VX-500, v3, v4, v5, and v6
FDA device enforcement

Medfusion Syringe Pump, Model 3500-0600-XX: a) 0600-00; b) 0600-01; c) 0600-50; d) 0600-51; e) 0600-82; f) 0600-249; software versions: v3, v4, v5, and v6

A force sensor in the occlusion detector may drift out of calibration leading to increased occlusion detection times, false occlusion alarms, or a System Failure Alarm. If the force sensor calibration shift is large enough, the pump will display a System Fa...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the country of Canada.

Official source for Medfusion Syringe Pump, Model 3500-0600-XX: a) 0600-00; b) 0600-01; c) 0600-50; d) 0600-51; e) 06...
FDA device enforcement

Medfusion Syringe Pump, Model 4000-XXXX-XX: a) 0100-50, b) 0100-249, c) 0101-50, d) 0101-51, e) 0101-78, f) 0101-249, g) 0105-51, h) 0105-78, i) 0105-249, j) 0106-00, k) 0106-01...

A force sensor in the occlusion detector may drift out of calibration leading to increased occlusion detection times, false occlusion alarms, or a System Failure Alarm. If the force sensor calibration shift is large enough, the pump will display a System Fa...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the country of Canada.

Official source for Medfusion Syringe Pump, Model 4000-XXXX-XX: a) 0100-50, b) 0100-249, c) 0101-50, d) 0101-51, e) 0...
FDA device enforcement

Medfusion PLUNGER FLOAT PLATE, Component Number G6000069, used on infusion pump

A force sensor in the occlusion detector may drift out of calibration leading to increased occlusion detection times, false occlusion alarms, or a System Failure Alarm. If the force sensor calibration shift is large enough, the pump will display a System Fa...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the country of Canada.

Official source for Medfusion PLUNGER FLOAT PLATE, Component Number G6000069, used on infusion pump
FDA device enforcement

Medfusion ASSEMBLY PLUNGER CASE LEFT 1/EA, Component Number G6001587, used on infusion pump

A force sensor in the occlusion detector may drift out of calibration leading to increased occlusion detection times, false occlusion alarms, or a System Failure Alarm. If the force sensor calibration shift is large enough, the pump will display a System Fa...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the country of Canada.

Official source for Medfusion ASSEMBLY PLUNGER CASE LEFT 1/EA, Component Number G6001587, used on infusion pump
FDA device enforcement

DuraLife Autoclavable Double Swivel Elbow, REF 60-0010 (UltraSet Product Code 66-1991); Swivel Elbow with Suction Port for use with breathing circuits

DuraLife devices instructions for cleaning and sterilization have been determined to be inadequate. Specifically, for the DuraLife product, the IFU does not provide the cleaning method, duration of autoclave cycle or specify the number of cycles the product...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Canada, New Zealand, and Hong Kong.

Official source for DuraLife Autoclavable Double Swivel Elbow, REF 60-0010 (UltraSet Product Code 66-1991); Swivel El...
FDA device enforcement

DuraLife Autoclavable Silicone Extension Tube, REF 60-1502 (UltraSet Product Code 66-2509); Non-Ported Double Swivel Elbow for use with breathing circuits

DuraLife devices instructions for cleaning and sterilization have been determined to be inadequate. Specifically, for the DuraLife product, the IFU does not provide the cleaning method, duration of autoclave cycle or specify the number of cycles the product...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Canada, New Zealand, and Hong Kong.

Official source for DuraLife Autoclavable Silicone Extension Tube, REF 60-1502 (UltraSet Product Code 66-2509); Non-P...
FDA device enforcement

DuraLife Autoclavable Silicone Extension Tube, REF 60-1510 (UltraSet Product Code 66-2505); Swivel Elbow with Suction Port for use with breathing circuits

DuraLife devices instructions for cleaning and sterilization have been determined to be inadequate. Specifically, for the DuraLife product, the IFU does not provide the cleaning method, duration of autoclave cycle or specify the number of cycles the product...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Canada, New Zealand, and Hong Kong.

Official source for DuraLife Autoclavable Silicone Extension Tube, REF 60-1510 (UltraSet Product Code 66-2505); Swive...
FDA device enforcement

CADD Administration Set, List Numbers: a) REF 21-7346-24; b) REF 21-7363-24

There are two potential issues with CADD Infusion System Infusion Sets: (1) Lack of Delivery or Underdelivery related to Tubing Occlusion and (2) False "No Disposable Attached (NDA)" Alarms

Class IReported ongoing

Distribution: Worldwide Distribution: US (nationwide) and OUS countries of: CH, CA, ES, FR, AT, BE, IT, DK, NO, EE, DE, NL, AU, GB, SE, PL, SA, IN, IL, PT, CZ, IE, NZ, LU, SI, ZA, BM, SG, MQ,...

Official source for CADD Administration Set, List Numbers: a) REF 21-7346-24; b) REF 21-7363-24
FDA device enforcement

CADD Administration Set, List Numbers: 1) REF 21-7300-24, 2) REF 21-7301-24 3) REF 21-7302-24, 4) REF 21-7321-24, 5) REF 21-7322-24, 6) REF 21-7323-24, 7) REF 21-7324-24, 8) REF...

There are two potential issues with CADD Infusion System Infusion Sets: (1) Lack of Delivery or Underdelivery related to Tubing Occlusion and (2) False "No Disposable Attached (NDA)" Alarms

Class IReported ongoing

Distribution: Worldwide Distribution: US (nationwide) and OUS countries of: CH, CA, ES, FR, AT, BE, IT, DK, NO, EE, DE, NL, AU, GB, SE, PL, SA, IN, IL, PT, CZ, IE, NZ, LU, SI, ZA, BM, SG, MQ,...

Official source for CADD Administration Set, List Numbers: 1) REF 21-7300-24, 2) REF 21-7301-24 3) REF 21-7302-24, 4)...
FDA device enforcement

CADD Administration Set, List Numbers: a) REF 21-7308-24, b) REF 21-7309-24, c) REF 21-7310-24, d) REF 21-7308-24JP, e) REF 21-7309-24JP. Computerized Ambulatory Delivery Device

There are two potential issues with CADD Infusion System Infusion Sets: (1) Lack of Delivery or Underdelivery related to Tubing Occlusion and (2) False "No Disposable Attached (NDA)" Alarms

Class IReported ongoing

Distribution: Worldwide Distribution: US (nationwide) and OUS countries of: CH, CA, ES, FR, AT, BE, IT, DK, NO, EE, DE, NL, AU, GB, SE, PL, SA, IN, IL, PT, CZ, IE, NZ, LU, SI, ZA, BM, SG, MQ,...

Official source for CADD Administration Set, List Numbers: a) REF 21-7308-24, b) REF 21-7309-24, c) REF 21-7310-24, d...
FDA device enforcement

CADD Administration Set, List Numbers: a) REF 21-7600-24, b) REF 21-7609-24, c) REF 21-7624-24, d) REF 21-7649-24, e) REF 21-7600-24JP, f) REF 21-7609-24JP, g) REF 21-7624-24JP,...

There are two potential issues with CADD Infusion System Infusion Sets: (1) Lack of Delivery or Underdelivery related to Tubing Occlusion and (2) False "No Disposable Attached (NDA)" Alarms

Class IReported ongoing

Distribution: Worldwide Distribution: US (nationwide) and OUS countries of: CH, CA, ES, FR, AT, BE, IT, DK, NO, EE, DE, NL, AU, GB, SE, PL, SA, IN, IL, PT, CZ, IE, NZ, LU, SI, ZA, BM, SG, MQ,...

Official source for CADD Administration Set, List Numbers: a) REF 21-7600-24, b) REF 21-7609-24, c) REF 21-7624-24, d...
FDA device enforcement

CADD Administration Set, List Numbers: 1) REF 21-7300-24, 2) REF 21-7301-24, 3) REF 21-7302-24, 4) REF 21-7308-24, 5) REF 21-7309-24, 6) REF 21-7310-24, 7) REF 21-7322-24, 8) RE...

There are two potential issues with CADD Infusion System Infusion Sets: (1) Lack of Delivery or Underdelivery related to Tubing Occlusion and (2) False "No Disposable Attached (NDA)" Alarms

Class IReported ongoing

Distribution: Worldwide Distribution: US (nationwide) and OUS countries of: CH, CA, ES, FR, AT, BE, IT, DK, NO, EE, DE, NL, AU, GB, SE, PL, SA, IN, IL, PT, CZ, IE, NZ, LU, SI, ZA, BM, SG, MQ,...

Official source for CADD Administration Set, List Numbers: 1) REF 21-7300-24, 2) REF 21-7301-24, 3) REF 21-7302-24, 4...
FDA device enforcement

Medfusion Model 4000 Syringe Infusion Pump, with software versions V1.6.0 or V1.6.1 that have implemented the Smart Pump Programming (SPP) feature in a clinical treatment enviro...

Potential for delay in sending Infusion Start Messages to third-party Electronic Health Record systems which may affect infusion auto-programming on the pump and documentation

Class IIReported ongoing

Distribution: US distribution to states of: Indiana, Texas and Virginia

Official source for Medfusion Model 4000 Syringe Infusion Pump, with software versions V1.6.0 or V1.6.1 that have imp...
FDA device enforcement

Medfusion Syringe Pump Model: Model 4000-0100-50, Model 4000-0101-249, Model 4000-0101-50, Model 4000-0101-51, Model 4000-0101-78, Model 4000-0105-249, Model 4000-0105-51, Model...

Multiple issues with the potential for interruption of therapy or over-infusion: 1. Primary Audible Alarm (PAA), 2. Unanticipated Depleted Battery Alarms, 3. Time Base Alarm, 4. Intermittent Volume Over Time (IVOT) - Infusion Continues after System Failure,...

Class IReported ongoing

Distribution: Worldwide distribution. US nationwide including US Virgin Islands, Countries of: AE, AL, AT, AU, BE, BM, BN, BS, CA, CH, CL, CY, CZ, DE, DK, DO, ES, FI, FR, GB, HK, HU, ID, IE, ...

Official source for Medfusion Syringe Pump Model: Model 4000-0100-50, Model 4000-0101-249, Model 4000-0101-50, Model ...
FDA device enforcement

Medfusion Syringe Pump Models: 3500, 3500-0600-00, 3500-0600-01, 3500-0600-249, 3500-0600-50, 3500-0600-51, 3500-0600-82, 3500-306, 3500-402, 3500-414, 3500-415, 3500-500, 3500B...

Multiple issues with the potential for interruption of therapy or over-infusion: 1. Primary Audible Alarm (PAA), 2. Unanticipated Depleted Battery Alarms, 3. Time Base Alarm, 4. Intermittent Volume Over Time (IVOT) - Infusion Continues after System Failure,...

Class IReported ongoing

Distribution: Worldwide distribution. US nationwide including US Virgin Islands, Countries of: AE, AL, AT, AU, BE, BM, BN, BS, CA, CH, CL, CY, CZ, DE, DK, DO, ES, FI, FR, GB, HK, HU, ID, IE, ...

Official source for Medfusion Syringe Pump Models: 3500, 3500-0600-00, 3500-0600-01, 3500-0600-249, 3500-0600-50, 350...

Reasons in these listed records

  • A force sensor in the occlusion detector may drift out of calibration leading to increased occlusion detection times, false occlusion alarms, or a System Failure Alarm. If the force sensor calibration shift is large enough, the pump will display a System Failure Alarm (includi...
  • Smiths Medical identified a labeling error in which the barcode (UDI) was not correctly reflecting the expiration date of the product
  • DuraLife devices instructions for cleaning and sterilization have been determined to be inadequate. Specifically, for the DuraLife product, the IFU does not provide the cleaning method, duration of autoclave cycle or specify the number of cycles the product can be reused on a ...
  • There are two potential issues with CADD Infusion System Infusion Sets: (1) Lack of Delivery or Underdelivery related to Tubing Occlusion and (2) False "No Disposable Attached (NDA)" Alarms

These examples come from this result page and are not a ranking of every recall reason.

Checking Smiths Medical ASD Inc.

Is there a recall on smiths medical asd inc. right now?

This archive contains 138 records, including 138 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.