CADD Administration Set, List Numbers: 1) REF 21-7300-24, 2) REF 21-7301-24 3) REF 21-7302-24, 4) REF 21-7321-24, 5) ... — FDA record Z-0963-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 9, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
CADD Administration Set, List Numbers: 1) REF 21-7300-24, 2) REF 21-7301-24 3) REF 21-7302-24, 4) REF 21-7321-24, 5) REF 21-7322-24, 6) REF 21-7323-24, 7) REF 21-7324-24, 8) REF 21-7333-24, 9) REF 21-7336-24, 10) REF 21-7339-24, 11) REF 21-7343-24, 12) REF 21-7349-24, 13) REF 21-7359-24, 14) REF 21-7383-24, 15) REF 21-7390-24, 16) REF 21-7391-24, 17) REF 21-7394-24, 18) REF 21-7395-24, 19) REF 21-7301-24JP, 20) REF 21-7302-24JP, 21) REF 21-7321-24JP, 22) REF 21-7322-24JP, 23) REF 21-7359-24JP, 24) REF 21-7394-24JP
Reason
There are two potential issues with CADD Infusion System Infusion Sets: (1) Lack of Delivery or Underdelivery related to Tubing Occlusion and (2) False "No Disposable Attached (NDA)" Alarms
Classification
Class I
Recalling firm
Smiths Medical ASD Inc.
Quantity
45285351 units
Agency-reported status
Ongoing
Source record identifier
Z-0963-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicability1) REF 21-7300-24, UDI/DI 30610586027196, Lot numbers: 3630772 through 4334107; 2) REF 21-7301-24, UDI/DI 30610586027219, Lot numbers: 3630747 through 4334119; 3) REF 21-7302-24, UDI/DI 30610586027233, Lot numbers: 3617363 through 4334138; 4) REF 21-7321-24, UDI/DI 30610586029633, Lot numbers: 3773534 through 4308545; 5) REF 21-7322-24, UDI/DI 30610586029640, Lot numbers: 3776375 through 4334324; 6) REF 21-7323-24, UDI/DI 061058603961, Lot numbers: 3776373 through 4334325; 7) REF 21-7324-24, UDI/DI 30610586029657, Lot numbers: 3773527 through 4334328; 8) REF 21-7333-24, UDI/DI 30610586043233, Lot numbers: 3776362 through 3984144; 9) REF 21-7336-24, UDI/DI 30610586027295, Lot numbers: 3776360 through 4025381; 10) REF 21-7339-24, UDI/DI 30610586027301, Lot numbers: 3780565 through 4009665; 11) REF 21-7343-24, UDI/DI 35019517191824, Lot numbers: 3965344 through 4334338; 12) REF 21-7349-24, UDI/DI 35019517184239, Lot numbers: 3926579 through 4339160; 13) REF 21-7359-24, UDI/DI 30610586027400, Lot numbers: 3776315 through 4334339; 14) REF 21-7383-24, UDI/DI 30610586027479, Lot numbers: 3780549 through 3971523; 15) REF 21-7390-24, UDI/DI 30610586027509, Lot numbers: 3780548 through 4334346; 16) REF 21-7391-24, UDI/DI 30610586027516, Lot numbers: 3773276 through 4334351; 17) REF 21-7394-24, UDI/DI 30610586027530, Lot numbers: 3774739 through 4339178; 18) REF 21-7395-24, UDI/DI 30610586027547, Lot numbers: 3808536 through 4308562 19) REF 21-7301-24JP, UDI/DI N/A , Lot numbers: 3630747 through 4334113; 20) REF 21-7302-24JP, UDI/DI N/A , Lot numbers: 3617363 through 4329630; 21) REF 21-7321-24JP, UDI/DI N/A , Lot numbers: 3773534 through 4308545; 22) REF 21-7322-24JP, UDI/DI N/A , Lot numbers: 3776375 through 4334324; 23) REF 21-7359-24JP, UDI/DI N/A , Lot numbers: 3776315 through 4334339; 24) REF 21-7394-24JP, UDI/DI N/A , Lot numbers: 3774739 through 4339178.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide Distribution: US (nationwide) and OUS countries of: CH, CA, ES, FR, AT, BE, IT, DK, NO, EE, DE, NL, AU, GB, SE, PL, SA, IN, IL, PT, CZ, IE, NZ, LU, SI, ZA, BM, SG, MQ, CL, LB, MY, HU, JP, TW, and CN

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 8, 2023
Recall initiation date
Dec 9, 2022
Center classification date
Jan 30, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records