Outset Medical, Inc. recall records

5 reported ongoing

5 records in this archive. Latest source date Aug 1, 2024.

FDA enforcement status can lag behind the recall lifecycle. An empty search or a reported termination is not a personal product-safety determination.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

Records and reported status

FDA device enforcement

Tablo Console, REF: PN-0003000, PN-0006000, part of the Tablo Hemodialysis System. The Tablo Hemodialysis System is indicated for use in patients with acute and/or chronic renal...

Blood leak sensor, a component of hemodialysis device, may show decreased sensitivity, at the lowest level, specified as 0.35 mL/min blood leak rate, which could delay triggering the ALARM_DIALYZER_BLOOD_LEAK alarm, if blood loss, less-than 300ml, occurs, t...

Class IIReported ongoing

Distribution: US: CA, NY, IL, MO, KY, AZ, PA, FL, GA, AK, NM, VA, NC, SC, OK, AL, MA, OH, MI, WA, ID, TX, TN, WY, IA, LA, NE, IN, AR, NV, WV, CO, NJ, MD, KS, UT, CT, MN, MS, WI, HI, DE, VT, M...

Official source for Tablo Console, REF: PN-0003000, PN-0006000, part of the Tablo Hemodialysis System. The Tablo Hemo...
FDA device enforcement

TabloCart with Prefiltration Drawer PN-0006813, with power cords: PN-0005746 (10-foot), PN-0006578 (15-foot), PN-0005747 (20-foot)10-, 15- and 20-foot cords respectively), used ...

A cart with prefiltration drawer, an optional hemodialysis system accessory, has a cord with a female connector that has internal contact pins that may become damaged that could create localized heat, sparks, smoke, melting or charring (thermal event) of th...

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of FL, CA, NC, GA, MD, DE, AK, OK, MA, WV, IN, KY, AZ, LA, NE, KS, NJ, CT, NM, CO, SC, TX, PA, NY, OH.

Official source for TabloCart with Prefiltration Drawer PN-0006813, with power cords: PN-0005746 (10-foot), PN-000657...
FDA device enforcement

Tablo Hemodialysis System, REF: PN-0003000, PN-0006000 and PN-0007001

Hemodialysis System, pre and post dialyzer peroxide-cured silicone tubing, with less than 336 hours usage, will be replaced because tubing may contain non-dioxin-like (NDL) polychlorinated biphenyl acids (PCBAs), that have not been flushed out over time wit...

Class IReported ongoing

Distribution: US:FL, CA, NM, SC, AZ, TN, NC, TX, IL, MD, IN, DE, AL, MI, AK, KY, LA, ID, OK, IA, VA, GA, MA, NE, KS, WI, PA, OH, WA, MT, RI, MN, MO, AR, NY, CO, WV, SD, OR, UT, WY, DC. OUS: AE

Official source for Tablo Hemodialysis System, REF: PN-0003000, PN-0006000 and PN-0007001
FDA device enforcement

Outset Tablo Model Numbers (REF): PN-0006000 PN-0007001 Tablo is a next generation, self-contained hemodialysis system (Hemodialysis System or System), intended for acute, home ...

An observed trend of high conductivity dialysate alarms in a hemodialysis system as a result of introducing a software tool into the manufacturing process.

Class IIReported ongoing

Distribution: U.S. Nationwide distribution including in the states of AK, AL, AR, CA, CO, FL, GA, HI, IA, ID, IL, IN, MI, MS, MO, MT, NE, NV, NJ, NM, NY, NC, OH, OK, OR, PA, SC, TN, TX, UT, V...

Official source for Outset Tablo Model Numbers (REF): PN-0006000 PN-0007001 Tablo is a next generation, self-containe...
FDA device enforcement

Outset Tablo Console Model Number: PN-0003000 Model Number: PN-0006000

Due to a component in the hemodialysis console there is the possibility of heat-related damage that may occur with the device.

Class IIReported ongoing

Distribution: U.S. Nationwide distribution in the states of AL, AK, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MS, MO, MT, NE, NV, NJ, NY, NM, NC, ND, OH,...

Official source for Outset Tablo Console Model Number: PN-0003000 Model Number: PN-0006000

Reasons in these listed records

  • Blood leak sensor, a component of hemodialysis device, may show decreased sensitivity, at the lowest level, specified as 0.35 mL/min blood leak rate, which could delay triggering the ALARM_DIALYZER_BLOOD_LEAK alarm, if blood loss, less-than 300ml, occurs, the patient may exper...
  • A cart with prefiltration drawer, an optional hemodialysis system accessory, has a cord with a female connector that has internal contact pins that may become damaged that could create localized heat, sparks, smoke, melting or charring (thermal event) of the female connector o...
  • Hemodialysis System, pre and post dialyzer peroxide-cured silicone tubing, with less than 336 hours usage, will be replaced because tubing may contain non-dioxin-like (NDL) polychlorinated biphenyl acids (PCBAs), that have not been flushed out over time with usage. Exposure to...
  • An observed trend of high conductivity dialysate alarms in a hemodialysis system as a result of introducing a software tool into the manufacturing process.

These examples come from this result page and are not a ranking of every recall reason.

Checking Outset Medical, Inc.

Is there a recall on outset medical, inc. right now?

This archive contains 5 records, including 5 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking