FDA device enforcement

Medtronic 1x4 Pocket Adaptor Kit for Deep Brain Stimulation, Model 64001; Stimulator, electrical, implanted, for parkinsonian tremor

A limited number of Pocket Adaptor Kits, Model 64001 and 64002, are labeled with incorrect Use-By-Dates.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and The countries of Austria, France, Germany, Italy, Japan, Latvia, Netherlands, Poland, Spain, Sweden, Switzerland, United Kingdom.

Official source for Medtronic 1x4 Pocket Adaptor Kit for Deep Brain Stimulation, Model 64001; Stimulator, electrical,...
FDA device enforcement

Medtronic 2x4 Pocket Adaptor Kit for Deep Brain Stimulation, Model 64002; Stimulator, electrical, implanted, for parkinsonian tremor

A limited number of Pocket Adaptor Kits, Model 64001 and 64002, are labeled with incorrect Use-By-Dates.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and The countries of Austria, France, Germany, Italy, Japan, Latvia, Netherlands, Poland, Spain, Sweden, Switzerland, United Kingdom.

Official source for Medtronic 2x4 Pocket Adaptor Kit for Deep Brain Stimulation, Model 64002; Stimulator, electrical,...
FDA device enforcement

Medtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer Software Application, loaded on the following devices: 1. PROG CT900A CLINICIAN TABLET US, Model Number CT900...

Software issue: When programming the SynchroMed Pump with the A810 CP App in Flex Infusion Mode, the infusion steps should automatically be arranged in consecutive order, from earliest to latest. However, in rare cases, the steps may be displayed out of ord...

Class IIReported ongoing

Distribution: Worldwide distribution.

Official source for Medtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer Software Application, loaded ...
FDA device enforcement

A71200 Vanta" Clinician Programmer Application (CP App) version 2.0.2465 and 2.0.2683, Product number A71200. Software is deployed and operated on Samsung tablets (CFN CT900).

Complaints received that Vanta A71200 CP App does not function as intended during use. There is a potential for prolonged or postponed surgical procedure.

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution.

Official source for A71200 Vanta" Clinician Programmer Application (CP App) version 2.0.2465 and 2.0.2683, Product nu...
FDA device enforcement

Restore Clinician Programmer Application (CP App), Model A71100 used with the following programmers: CT900A PROG CT900A CLINICIAN TABLET US CT900B PROG CT900B CLINICIAN TABLET E...

Software issue where a Device Reset message displayed on the app was unable to be cleared. In rare cases, this resulted in an inability to resume therapy, and the patient experienced a recurrence of underlying pain symptoms. Resolution would require surgica...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Argentina, Aruba, Australia, Austria, Belgium, Bosnia And Herzegovina, Brazil, Canada, Canary Islands, Chile, China, ...

Official source for Restore Clinician Programmer Application (CP App), Model A71100 used with the following programme...
FDA device enforcement

Enhanced Verify Evaluation Handset (CFN HH90130FA)

Evaluation handsets may not be able to communicate with the neurostimulator due to handsets having a faster processing speed than prior handsets. An inability to connect may cause the therapy to stop being delivered and a system error will be displayed. Pat...

Class IIReported ongoing

Distribution: US Nationwide distribution including in the states of AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, LA, MA, MD, MI, MN, MS, NC, NE, NJ, NM, NY, OH, OR, PA, TX, WA, WI.

Official source for Enhanced Verify Evaluation Handset (CFN HH90130FA)
FDA device enforcement

Stimulation RC Clinician Programmer Application, Model A71400, when used with the Inceptiv (Model 977119) and Intellis Pro (Model 977118) neurostimulators

There is a software issue that can permanently disable communication with an implantable neurostimulator (INS) when a specific intraoperative programming sequence is used.

Class IIReported ongoing

Distribution: Worldwide Distribution: US (Nationwide) and OUS (Global) to countries of: Austria, Belgium, Bulgaria, Canary Islands, Cyprus, Croatia, Czech Republic, Denmark, Finland, France, ...

Official source for Stimulation RC Clinician Programmer Application, Model A71400, when used with the Inceptiv (Model...
FDA device enforcement

A820 myPTM Software Application associated with Medtronic SynchroMed Pump and Infusion System.

Product complaints were received describing the A820 myPTM app taking longer than expected for patients to interact with their implantable pump.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canary Islands, Czech Republic, France, Germany, Greece, Ireland, Italy, Netherlands, Norway, Portu...

Official source for A820 myPTM Software Application associated with Medtronic SynchroMed Pump and Infusion System.
FDA device enforcement

Medtronic Percept RC Implantable Neurostimulators (INS), Model B35300.

A limited number of Percept RC Implantable Neurostimulators (INS), Model B35300 will not be able to communicate with the A610 DBS Clinician Programmer Application upon use and an "Invalid Device" message will appear.

Class IIReported ongoing

Distribution: International distribution to the countries of Germany, UK, and Israel.

Official source for Medtronic Percept RC Implantable Neurostimulators (INS), Model B35300.
FDA device enforcement

Medtronic Recharger Kits, RS6230, RS7230, and RS5220A, containing components WR9230 and WR9220A Recharger for rechargeable implantable neurostimulators. The Rechargers are used ...

The Wireless Rechargers in the kits may be unable to enter a recharging session upon first use.

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the states of AL, AZ, CA, FL, GA, ID, MD, MI, MN, MO, NC, NJ, OH, PA, SC, TX, VA, and WA. The countries of Austria, Belgium, Denmark, F...

Official source for Medtronic Recharger Kits, RS6230, RS7230, and RS5220A, containing components WR9230 and WR9220A R...
FDA device enforcement

Medtronic Handset with Communicator, Model TH90Q01, InterStim Smart Programmer and Communicator Kits, containing the Model HH90 Handset and the TM90 Communicator.

A subset of the handsets within the kit may not be able to complete the pairing process with the communicator upon initial setup.

Class IIReported ongoing

Distribution: US Nationwide distribution.

Official source for Medtronic Handset with Communicator, Model TH90Q01, InterStim Smart Programmer and Communicator K...
FDA device enforcement

A610 Clinician Programmer (CP) Software Application (app) with the following versions: Version 2.0.x (2.0.4584, 2.0.4594, 2.0.4605, 2.0.4630, and 2.0.4648), Version 3.0.x (3.0.1...

Patients implanted with a pocket adaptor (Model 64001 and/or 64002) are limited to HEAD ONLY MRI eligibility.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Canada, Canary Islands, China, Colombia, Costa Rica...

Official source for A610 Clinician Programmer (CP) Software Application (app) with the following versions: Version 2....
FDA device enforcement

Medtronic Ascenda Intrathecal Catheter, Models: a) 8780 (114 cm); b) 8781 (140 cm); c) 8784 (catheter pump revision kit)

Design update to the Ascenda Intrathecal Catheter Models 8780, 8781, and 8784. The intent of the design update is to reduce the potential for tissue growth into the Ascenda catheter connector which may potentially lead to catheter occlusion.

Class IIReported ongoing

Distribution: Worldwide Distribution: US (nationwide) and OUS (foreign) countries of: Argentina, Aruba, Australia, Austria, Belgium, Brazil, Canada, Canary Islands, Chile, Colombia, Cyprus, C...

Official source for Medtronic Ascenda Intrathecal Catheter, Models: a) 8780 (114 cm); b) 8781 (140 cm); c) 8784 (cath...

Reasons in these listed records

  • A limited number of Pocket Adaptor Kits, Model 64001 and 64002, are labeled with incorrect Use-By-Dates.
  • Software issue: When programming the SynchroMed Pump with the A810 CP App in Flex Infusion Mode, the infusion steps should automatically be arranged in consecutive order, from earliest to latest. However, in rare cases, the steps may be displayed out of order. If this occurs, ...
  • Complaints received that Vanta A71200 CP App does not function as intended during use. There is a potential for prolonged or postponed surgical procedure.
  • Software issue where a Device Reset message displayed on the app was unable to be cleared. In rare cases, this resulted in an inability to resume therapy, and the patient experienced a recurrence of underlying pain symptoms. Resolution would require surgical replacement of the...

These examples come from this result page and are not a ranking of every recall reason.

Checking Medtronic Neuromodulation

Is there a recall on medtronic neuromodulation right now?

This archive contains 45 records, including 43 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.