Restore Clinician Programmer Application (CP App), Model A71100 used with the following programmers: CT900A PROG CT90... — FDA record Z-0996-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 20, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Restore Clinician Programmer Application (CP App), Model A71100 used with the following programmers: CT900A PROG CT900A CLINICIAN TABLET US CT900B PROG CT900B CLINICIAN TABLET EU -UK CT900C PROG CT900C CLINICIAN TABLET UK CT900D PROG CT900D CLINICIAN TABLET GLOBAL CT900E PROG CT900E CLINICIAN TABLET GLBL SMSNG CT900F PROG CT900F CLINICIAN TABLET GLBL SMSNG
Reason
Software issue where a Device Reset message displayed on the app was unable to be cleared. In rare cases, this resulted in an inability to resume therapy, and the patient experienced a recurrence of underlying pain symptoms. Resolution would require surgical replacement of the INS.
Classification
Class II
Recalling firm
Medtronic Neuromodulation
Quantity
7667 devices
Agency-reported status
Ongoing
Source record identifier
Z-0996-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityCP App: UDI-DI 00763000273668 all software versions prior to v.1.0.4489 are affected The implanted devices are not affected.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide and the countries of Argentina, Aruba, Australia, Austria, Belgium, Bosnia And Herzegovina, Brazil, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, Germany, Greece, Guatemala, Hong Kong, Hungary, Iceland, Indonesia, Ireland, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Malta, Mauritius, Mexico, Namibia, Netherlands, New Zealand, Norway, Oman, Panama, Peru, Poland, Portugal, Qatar, Reunion, Russia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 31, 2025
Recall initiation date
Nov 20, 2025
Center classification date
Dec 23, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records