Maquet Cardiovascular, LLC recall records

35 reported ongoing

38 records in this archive. Latest source date Jun 2, 2026.

FDA enforcement status can lag behind the recall lifecycle. An empty search or a reported termination is not a personal product-safety determination.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

Records and reported status

FDA device enforcement

Vasoview Hemopro Endoscopic Vessel Harvesting System, model VH-3500.

Risk of polyimide particulate matter being present on the endoscopic harvesting vessel system. This issue may result in polyimide particulate matter being introduced into the arterial conduit for patients undergoing a coronary artery bypass graft (CABG) or ...

Class IIReported ongoing

Distribution: US: OUS: China, Brazil, Hong Kong.

Official source for Vasoview Hemopro Endoscopic Vessel Harvesting System, model VH-3500.
FDA device enforcement

Vasoview 6 Pro (VV6) - Endoscopic Vessel Harvesting System, Model VH-2400 printed on outer label, inner label and IFU, outer box artwork for Vasoview Hemopro 2 (HP2).

Device packaged inside an incorrect outer box; displaying artwork & illustrations for a different device.

Class IIReported ongoing

Distribution: US (AL, AR, CA, FL, GA, IN, KY, MI, NJ, OK, PA, TN, TX) and Canada.

Official source for Vasoview 6 Pro (VV6) - Endoscopic Vessel Harvesting System, Model VH-2400 printed on outer label,...
FDA device enforcement

Maquet Acrobat-i Vacuum Stabilizer System Intended for use during performance of cardiac surgical procedures through a sternotomy incision approach. Product Code: OM-10000Z

Housing Mount Jaw of these devices breaking during normal use and may result in injury to the epicardial tissue or vasculature, or a coronary artery or previously placed bypass graft, or could result in procedural delay and/or conversion.

Class IIReported ongoing

Distribution: Nationwide Foreign: AT BE DE ES FR GB IT JP KR LU NL PL PT TR

Official source for Maquet Acrobat-i Vacuum Stabilizer System Intended for use during performance of cardiac surgical...
FDA device enforcement

ACCESSRAIL Platform (Standard Blade) Part Number SB-1000, and (Deep Blades) Part Number C-DB-1000. Component of a chest stabilization system for cardiac and chest surgery.

The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device component into the patient, and/or a proced...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Bangladesh, Brazil, Chile, Colombia, Iran, Jordan, Malaysia, Mexico, Netherlands, Singapore, Sudan, Taiwan, Thailand,...

Official source for ACCESSRAIL Platform (Standard Blade) Part Number SB-1000, and (Deep Blades) Part Number C-DB-1000...
FDA device enforcement

ACROBAT SUV Vacuum Off-Pump System, Part Number C-OM-9000S. Component of a chest stabilization system for cardiac and chest surgery.

The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device component into the patient, and/or a proced...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Bangladesh, Brazil, Chile, Colombia, Iran, Jordan, Malaysia, Mexico, Netherlands, Singapore, Sudan, Taiwan, Thailand,...

Official source for ACROBAT SUV Vacuum Off-Pump System, Part Number C-OM-9000S. Component of a chest stabilization sy...
FDA device enforcement

ACROBAT V Vacuum Off-Pump System, Part Number C-OM-9100S. Component of a chest stabilization system for cardiac and chest surgery.

The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device component into the patient, and/or a proced...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Bangladesh, Brazil, Chile, Colombia, Iran, Jordan, Malaysia, Mexico, Netherlands, Singapore, Sudan, Taiwan, Thailand,...

Official source for ACROBAT V Vacuum Off-Pump System, Part Number C-OM-9100S. Component of a chest stabilization syst...
FDA device enforcement

ACROBAT-i Vacuum Stabilizer System, Part Number C-OM-10000. Component of a chest stabilization system for cardiac and chest surgery.

The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device component into the patient, and/or a proced...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Bangladesh, Brazil, Chile, Colombia, Iran, Jordan, Malaysia, Mexico, Netherlands, Singapore, Sudan, Taiwan, Thailand,...

Official source for ACROBAT-i Vacuum Stabilizer System, Part Number C-OM-10000. Component of a chest stabilization sy...
FDA device enforcement

Fusion Bioline Vascular Graft, Part number M00201503046B0

One (1) Fusion Bioline Vascular graft from lot 25162546, which failed guideline straightness testing, was intermingled with lot 25162549 and distributed to customers. Use of the affected graft may result in reduction of blood flow or unwanted tension causin...

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of FL, IN, LA, MI, MO, NC, NV, NY, PA, TN, TX, VA, WA.

Official source for Fusion Bioline Vascular Graft, Part number M00201503046B0
FDA device enforcement

INTERGARD Woven Vascular Graft is a woven polyester graft coated with a bovine collagen. It has an external velour surface and non-velour inner surface. The device is available ...

Mislabeled: Product labeled as Intergard Woven Straight graft REF IGW0032-30, D: 8 mm, L: 30 cm, contained an Intergard Woven Straight graft of D: 32 mm and L: 30 cm

Class IIReported terminated

Distribution: US Nationwide distribution in the state of SC.

Official source for INTERGARD Woven Vascular Graft is a woven polyester graft coated with a bovine collagen. It has a...
FDA device enforcement

Ultima Activator II Reusable Drive Mechanism, part # C-UA-5001. Labeled REF UA-5001.

Potential for corrosion on the pins of the Ultima Activator II Reusable Drive Mechanism. Harms may include allergic reaction, metal toxicity, and other delayed responses.

Class IIReported ongoing

Distribution: Worldwide - United States (AL, AZ, CA, FL, IL, KY, LA, MO, MT, NM, OH, PA, TX, WI), Germany, Republic of Korea (South Korea), Russia, Thailand, United Arab Emirates.

Official source for Ultima Activator II Reusable Drive Mechanism, part # C-UA-5001. Labeled REF UA-5001.
FDA device enforcement

HEMASHIELD GOLD KNITTED MICROVEL DOUBLE VELOUR VASCULAR GRAFT- D: 7 mm; L: 60 cm Indicated for use in the replacement or repair of arteries affected with aneurysmal or occlusive...

Outer labelling (HEMASHIELD GOLD Knitted Microvel Double Velour Bifurcated Vascular Graft, D: 16 x 8 mm; L: 40 cm) does not correspond to the (inner) sterile packaging labelling (HEMASHIELD GOLD Knitted Microvel Double Velour Straight Vascular Graft, D: 7 m...

Class IIReported terminated

Distribution: US Nationwide distribution in the state of MO.

Official source for HEMASHIELD GOLD KNITTED MICROVEL DOUBLE VELOUR VASCULAR GRAFT- D: 7 mm; L: 60 cm Indicated for us...
FDA device enforcement

Hemashield Gold Knitted Bifurcated, 18 mm x 9 mm; reference number M002020851890

There is a probable mislabeling of product. A Hemashield Platinum Woven Double Velour Bifurcated Vasuclar Graft, SN 1259735369, 24 mm x 12 mm, REF Number M00202166241P0 is likely packaged inside of the carton instead of a Hemashield Gold Knitted Microvel Do...

Class IIReported terminated

Distribution: US Nationwide distribution in the states of IL and NC.

Official source for Hemashield Gold Knitted Bifurcated, 18 mm x 9 mm; reference number M002020851890

Reasons in these listed records

  • Risk of polyimide particulate matter being present on the endoscopic harvesting vessel system. This issue may result in polyimide particulate matter being introduced into the arterial conduit for patients undergoing a coronary artery bypass graft (CABG) or peripheral bypass gr...
  • The Positioner Arm may not tighten or lock during normal use.
  • Device packaged inside an incorrect outer box; displaying artwork & illustrations for a different device.
  • Housing Mount Jaw of these devices breaking during normal use and may result in injury to the epicardial tissue or vasculature, or a coronary artery or previously placed bypass graft, or could result in procedural delay and/or conversion.

These examples come from this result page and are not a ranking of every recall reason.

Checking Maquet Cardiovascular, LLC

Is there a recall on maquet cardiovascular, llc right now?

This archive contains 38 records, including 35 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking