Ultima Activator II Reusable Drive Mechanism, part # C-UA-5001. Labeled REF UA-5001. — FDA record Z-0887-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Feb 9, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Ultima Activator II Reusable Drive Mechanism, part # C-UA-5001. Labeled REF UA-5001.
Reason
Potential for corrosion on the pins of the Ultima Activator II Reusable Drive Mechanism. Harms may include allergic reaction, metal toxicity, and other delayed responses.
Classification
Class II
Recalling firm
Maquet Cardiovascular, LLC
Quantity
60
Agency-reported status
Ongoing
Source record identifier
Z-0887-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI# 00607567700901, Lots (3): 25153700, 25157489, 25158616. Please note affected lot numbers (25153700, 25157489, 25158616) appear on the device package only. If the packaging has been discarded, the affected devices can be identified by Device Component Lot number (200501), which is etched onto all 3 device components on side of handle and on underside above each pin.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - United States (AL, AZ, CA, FL, IL, KY, LA, MO, MT, NM, OH, PA, TX, WI), Germany, Republic of Korea (South Korea), Russia, Thailand, United Arab Emirates.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Apr 13, 2022
Recall initiation date
Feb 9, 2022
Center classification date
Apr 6, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.