Records and reported status

FDA device enforcement

BD MAX System, BD MAX Instrument Version or Model: 441916 Catalog Number: 441916

BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may ...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bhutan, Bo...

Official source for BD MAX System, BD MAX Instrument Version or Model: 441916 Catalog Number: 441916
FDA device enforcement

BD Synapsys Informatics Solution (Laboratory Solution) Version or Model: 444150 Catalog Number: 444150

BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may ...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bhutan, Bo...

Official source for BD Synapsys Informatics Solution (Laboratory Solution) Version or Model: 444150 Catalog Number: 4...
FDA device enforcement

BD Phoenix M50 Automated Microbiology System Version or Model: 443624 Catalog Number: 443624

BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may ...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bhutan, Bo...

Official source for BD Phoenix M50 Automated Microbiology System Version or Model: 443624 Catalog Number: 443624
FDA device enforcement

BD BACTEC FX, Instrument Top, Packaged Version or Model: 441385 Catalog Number: 441385

BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may ...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bhutan, Bo...

Official source for BD BACTEC FX, Instrument Top, Packaged Version or Model: 441385 Catalog Number: 441385
FDA device enforcement

BD BACTEC FX40 Instrument Version or Model: 442296 Catalog Number: 442296

BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may ...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bhutan, Bo...

Official source for BD BACTEC FX40 Instrument Version or Model: 442296 Catalog Number: 442296
FDA device enforcement

BACTEC 9120 System Version or Model: 445570 Catalog Number: 445570

BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may ...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bhutan, Bo...

Official source for BACTEC 9120 System Version or Model: 445570 Catalog Number: 445570
FDA device enforcement

BACTEC 9240 System Version or Model: 445475 Catalog Number: 445475

BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may ...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bhutan, Bo...

Official source for BACTEC 9240 System Version or Model: 445475 Catalog Number: 445475
FDA device enforcement

BD Synapsys Informatics Solution with software versions: 4.20, 4.30, 5.10, 5.20 and 5.30- IVD standalone laboratory software solution providing data management and workflow mana...

Synapsys software versions 4.20, 4.30, 5.10, 5.20, or 5.30 and utilizing the in-line breakpoint feature, an error is occurring where BD Synapsys will incorrectly display breakpoints for organisms that do not have defined breakpoints in CLSI or EUCAST. This ...

Class IIReported ongoing

Distribution: Nationwide Foreign: Australia, New Zealand, Singapore, Hong Kong, Taiwan, Japan, Austria, Belgium, Denmark, France, Germany, Italy, Luxembourg, Netherlands, Norway, Spain, Swede...

Official source for BD Synapsys Informatics Solution with software versions: 4.20, 4.30, 5.10, 5.20 and 5.30- IVD sta...
FDA device enforcement

BD Phoenix M50 instrument -Intended for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically significant bacteria. The BD Phoenix System pr...

Escherichia coli (E. coli) misidentification for clinical samples that were tested on BD Phoenix" M50 instrument and BD Phoenix" 100, may lead to misdiagnosis and inappropriate treatment of infections caused by E. coli,

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia a...

Official source for BD Phoenix M50 instrument -Intended for the rapid identification (ID) and antimicrobial susceptib...
FDA device enforcement

BD Phoenix 100-Intended for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically significant bacteria. The BD Phoenix System provides rapid...

Escherichia coli (E. coli) misidentification for clinical samples that were tested on BD Phoenix" M50 instrument and BD Phoenix" 100, may lead to misdiagnosis and inappropriate treatment of infections caused by E. coli,

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia a...

Official source for BD Phoenix 100-Intended for the rapid identification (ID) and antimicrobial susceptibility testin...
FDA device enforcement

BD BACTEC MGIT 960 PZA Kit)- IVD qualitative procedure for susceptibility testing of Mycobacterium tuberculosis, from culture to pyrazinamide (PZA) Catalog Number: 245128

May intermittently produce falsely resistant results for pyrazinamide (PZA) during susceptibility testing of Mycobacterium tuberculosis isolates.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brunei, Brazil, Canada, China, Costa Rica, Dominican Republic, Guatemala, ...

Official source for BD BACTEC MGIT 960 PZA Kit)- IVD qualitative procedure for susceptibility testing of Mycobacteriu...
FDA device enforcement

BD Kiestra Urine Culture Application Powered by BD Synapsys Informatics Solution; Catalog (Ref): 444900;

There is a potential risk that the image colony count does not adequately reflect the actual plate colony count and may lead to inaccurate counts that would impact patient diagnosis and subsequent care.

Class IIReported ongoing

Distribution: US distribution to states of: CO, CT, IA, IL, KY, MD, MI, MN, NC, NJ, NY, OH, OK, SC, TN, TX and WA.

Official source for BD Kiestra Urine Culture Application Powered by BD Synapsys Informatics Solution; Catalog (Ref): ...
FDA device enforcement

BD SARS-CoV-2 Reagents for BD MAX" Systems. RT-PCR test to detect SARS-CoV-2 in nasal, nasopharyngeal, oropharyngeal swab specimens, and saliva specimens.

Assays may produce false negative results due to decreased activity of the enzyme utilized in manufacturing.

Class IIReported ongoing

Distribution: Domestic: Nationwide Distribution; Foreign: AU, BR, BE, CA, CO, & JP.

Official source for BD SARS-CoV-2 Reagents for BD MAX" Systems. RT-PCR test to detect SARS-CoV-2 in nasal, nasopharyn...
FDA device enforcement

BD Phoenix NMIC-306 Panel is a self-inoculating (when used with a BD Phoenix Automated Microbiology System) polystyrene panel (tray) containing control wells and substrate wells...

Products were found to have confirmed drug cross contamination prior to release may lead to incorrect, inadequate, or absent antibiotic treatment or delay in results

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of AR, BE, CO, IN, JP, PE, PY, SG, TW, UY.

Official source for BD Phoenix NMIC-306 Panel is a self-inoculating (when used with a BD Phoenix Automated Microbiolo...
FDA device enforcement

BD Phoenix Yeast ID Panel is a self-inoculating (when used with a BD Phoenix Automated Microbiology System) polystyrene panel (tray) containing control wells and substrate wells...

Esculin, a substrate used in other panels and normally in A16 well, was found in A17 well. A17 in yeast panels is supposed to be an empty well and is not a part of the normal ID evaluation. However, A17 is checked for fluorescence and if anything is found, ...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of AR, BE, CO, IN, JP, PE, PY, SG, TW, UY.

Official source for BD Phoenix Yeast ID Panel is a self-inoculating (when used with a BD Phoenix Automated Microbiolo...
FDA device enforcement

Sensi Disc Amoxicillin with Clavulanic Acid 20/10 ¿g, Catalog No. 231628 and 231629.¿ These discs are used for semi-quantitative in vitro susceptibility testing

There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, t...

Class IIReported ongoing

Distribution: Domestic distribution nationwide. International distribution worldwide.

Official source for Sensi Disc Amoxicillin with Clavulanic Acid 20/10 ¿g, Catalog No. 231628 and 231629.¿ These discs...
FDA device enforcement

BD BBL Sensi Disc Ampicillin - 10 ¿g, Catalog No. 230705 and 231264; 2 ¿g, Catalog No. 231263. Used for semi-quantitative in vitro susceptibility testing.

There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, t...

Class IIReported ongoing

Distribution: Domestic distribution nationwide. International distribution worldwide.

Official source for BD BBL Sensi Disc Ampicillin - 10 ¿g, Catalog No. 230705 and 231264; 2 ¿g, Catalog No. 231263. Us...
FDA device enforcement

BD BBL Sensi Disc Ampicillin with Sulbactam 10/10 ¿g, Catalog Number 231660. Used for semi-quantitative in vitro susceptibility testing.

There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, t...

Class IIReported ongoing

Distribution: Domestic distribution nationwide. International distribution worldwide.

Official source for BD BBL Sensi Disc Ampicillin with Sulbactam 10/10 ¿g, Catalog Number 231660. Used for semi-quanti...
FDA device enforcement

BD BBL Sensi Disc Augmentin - 3¿g, Catalog Number 291270¿used for semi-quantitative in vitro susceptibility testing.

There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, t...

Class IIReported ongoing

Distribution: Domestic distribution nationwide. International distribution worldwide.

Official source for BD BBL Sensi Disc Augmentin - 3¿g, Catalog Number 291270¿used for semi-quantitative in vitro susc...
FDA device enforcement

BD BBL Sensi Disc Azithromycin -15 ¿g, Catalog Number 231682¿used for semi-quantitative in vitro susceptibility testing.

There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, t...

Class IIReported ongoing

Distribution: Domestic distribution nationwide. International distribution worldwide.

Official source for BD BBL Sensi Disc Azithromycin -15 ¿g, Catalog Number 231682¿used for semi-quantitative in vitro ...

Reasons in these listed records

  • BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confident...
  • Synapsys software versions 4.20, 4.30, 5.10, 5.20, or 5.30 and utilizing the in-line breakpoint feature, an error is occurring where BD Synapsys will incorrectly display breakpoints for organisms that do not have defined breakpoints in CLSI or EUCAST. This occurs when the iden...
  • Escherichia coli (E. coli) misidentification for clinical samples that were tested on BD Phoenix" M50 instrument and BD Phoenix" 100, may lead to misdiagnosis and inappropriate treatment of infections caused by E. coli,
  • May intermittently produce falsely resistant results for pyrazinamide (PZA) during susceptibility testing of Mycobacterium tuberculosis isolates.

These examples come from this result page and are not a ranking of every recall reason.

Checking Becton Dickinson & Co.

Is there a recall on becton dickinson & co. right now?

This archive contains 73 records, including 72 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking