Recall archive / FDA device enforcement
BD Phoenix Yeast ID Panel is a self-inoculating (when used with a BD Phoenix Automated Microbiology System) polystyre... — FDA record Z-1399-2024
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Feb 29, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- BD Phoenix Yeast ID Panel is a self-inoculating (when used with a BD Phoenix Automated Microbiology System) polystyrene panel (tray) containing control wells and substrate wells with dried biochemical substrates that use traditional qualitative microbiology methods such as fermentation, oxidation, degradation, and hydrolysis in combination with chromogenic and fluorogenic substrates along with carbon sources for yeast identification.
- Reason
- Esculin, a substrate used in other panels and normally in A16 well, was found in A17 well. A17 in yeast panels is supposed to be an empty well and is not a part of the normal ID evaluation. However, A17 is checked for fluorescence and if anything is found, the system will abort the panel with the message Fluorescent interference has been detected. Esculin is expected to trigger this fluorescence and test cancellation.
- Classification
- Class II
- Recalling firm
- Becton Dickinson & Co.
- Quantity
- 6700 eaches
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1399-2024
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | UDI-DI: (10) 3312180 (17) 241030 (30)0025 (01) 30382904483167 Lot Number: 3312180 Exp. Date: 30-OCT-2024 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide distribution - US Nationwide and the countries of AR, BE, CO, IN, JP, PE, PY, SG, TW, UY.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Apr 10, 2024
- Recall initiation date
- Feb 29, 2024
- Center classification date
- Mar 29, 2024
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.