Baxter Healthcare Corporation recall records

273 reported ongoing

306 records in this archive. Latest source date Aug 17, 2026.

FDA enforcement status can lag behind the recall lifecycle. An empty search or a reported termination is not a personal product-safety determination.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

Records and reported status

FDA device enforcement

Baxter MiniCap Extended Life PD Transfer Set Twist Clamp - Extra Short, REF R5C4483

Cleaning products, containing, but not limited to, hydrogen peroxide, bleach, alcohol, or antiseptic agents and solvent, such as those containing acetone, toluene, xylene, or cyclohexanone may cause damage, such as leaking or cracking, to the MiniCap Extend...

Class IIReported ongoing

Distribution: Worldwide

Official source for Baxter MiniCap Extended Life PD Transfer Set Twist Clamp - Extra Short, REF R5C4483
FDA device enforcement

Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp - Extra Long, REF R5C4484

Cleaning products, containing, but not limited to, hydrogen peroxide, bleach, alcohol, or antiseptic agents and solvent, such as those containing acetone, toluene, xylene, or cyclohexanone may cause damage, such as leaking or cracking, to the MiniCap Extend...

Class IIReported ongoing

Distribution: Worldwide

Official source for Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp - Extra Long, REF R5C4484
FDA device enforcement

Baxter CAPD System Transfer Set (CAPD MiniCap Transfer Set II (24cm)), Code No. T5C4482

Cleaning products, containing, but not limited to, hydrogen peroxide, bleach, alcohol, or antiseptic agents and solvent, such as those containing acetone, toluene, xylene, or cyclohexanone may cause damage, such as leaking or cracking, to the MiniCap Extend...

Class IIReported ongoing

Distribution: Worldwide

Official source for Baxter CAPD System Transfer Set (CAPD MiniCap Transfer Set II (24cm)), Code No. T5C4482
FDA device enforcement

Baxter CAPD System Transfer Set (CAPD MiniCap Transfer Set II (34cm)), Code No. T5C4484

Cleaning products, containing, but not limited to, hydrogen peroxide, bleach, alcohol, or antiseptic agents and solvent, such as those containing acetone, toluene, xylene, or cyclohexanone may cause damage, such as leaking or cracking, to the MiniCap Extend...

Class IIReported ongoing

Distribution: Worldwide

Official source for Baxter CAPD System Transfer Set (CAPD MiniCap Transfer Set II (34cm)), Code No. T5C4484
FDA device enforcement

Baxter Spectrum IQ Infusion Pumps, Product code 3570009.

There is the potential for reduced or non-delivery of medication, in some cases without alerting the user via pump alarm. This may occur as a result of incorrect administration set setup and/or incomplete resolution of upstream occlusion alarms when using S...

Class IReported ongoing

Distribution: Worldwide Distribution. US nationwide Puerto Rico and St. Thomas. There was government/military distribution. Countries of Canada, Bahamas, Barbados, Bermuda, Guyana, Jamaica, a...

Official source for Baxter Spectrum IQ Infusion Pumps, Product code 3570009.
FDA device enforcement

Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 8), Product code 35700BAX2.

There is the potential for reduced or non-delivery of medication, in some cases without alerting the user via pump alarm. This may occur as a result of incorrect administration set setup and/or incomplete resolution of upstream occlusion alarms when using S...

Class IReported ongoing

Distribution: Worldwide Distribution. US nationwide Puerto Rico and St. Thomas. There was government/military distribution. Countries of Canada, Bahamas, Barbados, Bermuda, Guyana, Jamaica, a...

Official source for Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 8), Product code 35700BAX2.

Reasons in these listed records

  • Cleaning products, containing, but not limited to, hydrogen peroxide, bleach, alcohol, or antiseptic agents and solvent, such as those containing acetone, toluene, xylene, or cyclohexanone may cause damage, such as leaking or cracking, to the MiniCap Extended Life PD transfer ...
  • There is the potential for reduced or non-delivery of medication, in some cases without alerting the user via pump alarm. This may occur as a result of incorrect administration set setup and/or incomplete resolution of upstream occlusion alarms when using Spectrum V8 and Spect...

These examples come from this result page and are not a ranking of every recall reason.

Checking Baxter Healthcare Corporation

Is there a recall on baxter healthcare corporation right now?

This archive contains 306 records, including 273 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking