Baxter Spectrum IQ Infusion Pumps, Product code 3570009. — FDA record Z-0529-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 29, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Baxter Spectrum IQ Infusion Pumps, Product code 3570009.
Reason
There is the potential for reduced or non-delivery of medication, in some cases without alerting the user via pump alarm. This may occur as a result of incorrect administration set setup and/or incomplete resolution of upstream occlusion alarms when using Spectrum V8 and Spectrum IQ infusion pumps.
Classification
Class I
Recalling firm
Baxter Healthcare Corporation
Quantity
241,304 units
Agency-reported status
Ongoing
Source record identifier
Z-0529-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityAll serial numbers, GTIN 00085412610900.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide Distribution. US nationwide Puerto Rico and St. Thomas. There was government/military distribution. Countries of Canada, Bahamas, Barbados, Bermuda, Guyana, Jamaica, and Trinidad and Tobago.