FDA device enforcement

Baxter DISCPAC Syringe Tip Caps, Self-Righting, Luer Slip, Green: a) 25 pack, REF H93869025, b) 100 pack, REF H938674100, c) 100 pack, REF H93869100

There is a potential of the packaging not maintaining a sterile barrier for the tip caps.

Class IIReported ongoing

Distribution: Worldwide distribution: US (nationwide) distribution and OUS (foreign) : Australia, Finland, Germany, Spain and United Kingdom.

Official source for Baxter DISCPAC Syringe Tip Caps, Self-Righting, Luer Slip, Green: a) 25 pack, REF H93869025, b) 1...
FDA device enforcement

The Exacta-Mix 2400 compounding system, manufactured by Baxter Healthcare Corporation, is a compounding device used in the pharmacy to compound multiple source ingredients into ...

There is a potential for leaking valves on ports 1 to 4 in certain lots of valve sets.

Class IIReported ongoing

Distribution: Worldwide distribution: US Nationwide, Canada, Uruguay, Argentina, Brazil, Colombia, Dominican Republic, Panama, Puerto Rico, Hong Kong, Taiwan, Korea, United Kingdom, Austria, ...

Official source for The Exacta-Mix 2400 compounding system, manufactured by Baxter Healthcare Corporation, is a compo...
FDA device enforcement

Hill-Rom PRO+ 36" MRS Surface With X-Ray mattress, REF P7924A01; and REF P7924ARENT01, PRO+ MRS Rental Surface with X-Ray mattress.

The failure mode on the affected mattresses can cause a reduction in the performance of the Microclimate Management feature used to pull heat and moisture away from the patient resulting in an increased patient risk of pressure ulcer development.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, CT, GA, IA, IN, KY, LA, MD, MI, MN, MO, NC, NJ, NM, NM, NY, OH, OK, RI, SD, TN, TX, VT, and WI. ...

Official source for Hill-Rom PRO+ 36" MRS Surface With X-Ray mattress, REF P7924A01; and REF P7924ARENT01, PRO+ MRS R...
FDA device enforcement

Hill-Rom PRO+ 36" MRS Surface mattress, REF P7924A03.

The failure mode on the affected mattresses can cause a reduction in the performance of the Microclimate Management feature used to pull heat and moisture away from the patient resulting in an increased patient risk of pressure ulcer development.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, CT, GA, IA, IN, KY, LA, MD, MI, MN, MO, NC, NJ, NM, NM, NY, OH, OK, RI, SD, TN, TX, VT, and WI. ...

Official source for Hill-Rom PRO+ 36" MRS Surface mattress, REF P7924A03.
FDA device enforcement

Hill-Rom PRO+ MRS Surface For Versacare With X-Ray mattress, REF P3255A01.

The failure mode on the affected mattresses can cause a reduction in the performance of the Microclimate Management feature used to pull heat and moisture away from the patient resulting in an increased patient risk of pressure ulcer development.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, CT, GA, IA, IN, KY, LA, MD, MI, MN, MO, NC, NJ, NM, NM, NY, OH, OK, RI, SD, TN, TX, VT, and WI. ...

Official source for Hill-Rom PRO+ MRS Surface For Versacare With X-Ray mattress, REF P3255A01.
FDA device enforcement

Hill-Rom PRO+ MRS Surface For Versacare mattress, REF P3255A02.

The failure mode on the affected mattresses can cause a reduction in the performance of the Microclimate Management feature used to pull heat and moisture away from the patient resulting in an increased patient risk of pressure ulcer development.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, CT, GA, IA, IN, KY, LA, MD, MI, MN, MO, NC, NJ, NM, NM, NY, OH, OK, RI, SD, TN, TX, VT, and WI. ...

Official source for Hill-Rom PRO+ MRS Surface For Versacare mattress, REF P3255A02.
FDA device enforcement

Hill-Rom Centrella Pro+ 36" Surface With X-Ray mattress, REF P7923A01.

The failure mode on the affected mattresses can cause a reduction in the performance of the Microclimate Management feature used to pull heat and moisture away from the patient resulting in an increased patient risk of pressure ulcer development.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, CT, GA, IA, IN, KY, LA, MD, MI, MN, MO, NC, NJ, NM, NM, NY, OH, OK, RI, SD, TN, TX, VT, and WI. ...

Official source for Hill-Rom Centrella Pro+ 36" Surface With X-Ray mattress, REF P7923A01.
FDA device enforcement

Hill-Rom Centrella Pro+ 36" Surface mattress, REF P7923A03.

The failure mode on the affected mattresses can cause a reduction in the performance of the Microclimate Management feature used to pull heat and moisture away from the patient resulting in an increased patient risk of pressure ulcer development.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, CT, GA, IA, IN, KY, LA, MD, MI, MN, MO, NC, NJ, NM, NM, NY, OH, OK, RI, SD, TN, TX, VT, and WI. ...

Official source for Hill-Rom Centrella Pro+ 36" Surface mattress, REF P7923A03.
FDA device enforcement

Hill-Rom Centrella Pro+ 40" Surface mattress, REF P7923A04.

The failure mode on the affected mattresses can cause a reduction in the performance of the Microclimate Management feature used to pull heat and moisture away from the patient resulting in an increased patient risk of pressure ulcer development.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, CT, GA, IA, IN, KY, LA, MD, MI, MN, MO, NC, NJ, NM, NM, NY, OH, OK, RI, SD, TN, TX, VT, and WI. ...

Official source for Hill-Rom Centrella Pro+ 40" Surface mattress, REF P7923A04.
FDA device enforcement

ABACUS TPN (Total Parenteral Nutrition) Calculation software, Product codes: 8300-0167 (Abacus V3.1 CE), 8300-0168 (Abacus V3.1 SE), 8300-0169 (Abacus V3.1 ME), 8300-0191 (Abacu...

There is a potential risk of medication error when using specific Abacus Order Entry & Calculation software.

Class IReported terminated

Distribution: United States (including Puerto Rico), Argentina, Bermuda, Chile, Colombia, Dominican Republic, Panama, Canada, Vietnam, Hong Kong, Singapore, and EMEA

Official source for ABACUS TPN (Total Parenteral Nutrition) Calculation software, Product codes: 8300-0167 (Abacus V3...
FDA device enforcement

In-Line ventilator adaptor

There is a potential of reduced oxygenation or pneumothorax/barotrauma to occur when the Volara system is used with in-line ventilator in home care environment.

Class IReported terminated

Distribution: US Nationwide distribution in the states of TX, FL, VA, KY, CA, AZ, WI, IL, NC, OH, MI, MS, MO, MN, AL, TN, SD, OK, NV, NJ, IN, CO, GA, WA, PA, IA, NE, NM, SC, ND, KS, WV, OR.

Official source for In-Line ventilator adaptor
FDA device enforcement

In-Line ventilator adaptor

There is a potential of reduced oxygenation or pneumothorax/barotrauma to occur when the Volara system is used with in-line ventilator in home care environment.

Class IReported terminated

Distribution: US Nationwide distribution in the states of TX, FL, VA, KY, CA, AZ, WI, IL, NC, OH, MI, MS, MO, MN, AL, TN, SD, OK, NV, NJ, IN, CO, GA, WA, PA, IA, NE, NM, SC, ND, KS, WV, OR.

Official source for In-Line ventilator adaptor
FDA device enforcement

15 Liters Drain Bag

Certain Peritoneal Dialysis product codes do not have FDA regulatory clearance and were distributed within the United States in error.

Class IIReported terminated

Distribution: US Nationwide distribution in the states of CA, MO, MA, AZ, TX, FL, NY, MD, OH.

Official source for 15 Liters Drain Bag
FDA device enforcement

Effluent Sample Bag

Certain Peritoneal Dialysis product codes do not have FDA regulatory clearance and were distributed within the United States in error.

Class IIReported terminated

Distribution: US Nationwide distribution in the states of CA, MO, MA, AZ, TX, FL, NY, MD, OH.

Official source for Effluent Sample Bag

Reasons in these listed records

  • There is a potential of the packaging not maintaining a sterile barrier for the tip caps.
  • There is a potential for leaking valves on ports 1 to 4 in certain lots of valve sets.
  • The failure mode on the affected mattresses can cause a reduction in the performance of the Microclimate Management feature used to pull heat and moisture away from the patient resulting in an increased patient risk of pressure ulcer development.
  • There is a potential risk of medication error when using specific Abacus Order Entry & Calculation software.

These examples come from this result page and are not a ranking of every recall reason.

Checking Baxter Healthcare Corporation

Is there a recall on baxter healthcare corporation right now?

This archive contains 306 records, including 273 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking