NaviCare Nurse Call/Voalte Nurse Call — FDA record Z-1306-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 31, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
NaviCare Nurse Call/Voalte Nurse Call
Reason
An issue has been identified with Phillips (Emergin) and Longleaf non relay (Connexall, Vocera, Cerner) wireless integrations used with NaviCare/Voalte Nurse Call, software versions 3.9.100 through 3.9.300. Calls placed from a push button call device, such as a push button switch, call cord, or from the push buttons on a room audio station, will be canceled on the nurse call system when the call is answered at the wireless phone, regardless of the call priority.
Classification
Class II
Recalling firm
Baxter Healthcare Corporation
Quantity
283 installations
Agency-reported status
Ongoing
Source record identifier
Z-1306-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilitysoftware versions: 3.9.100, 3.9.101,3.9.102, 3.9.200, 3.9.201, and 3.9.300

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

nationwide and Canada

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jul 6, 2022
Recall initiation date
May 31, 2022
Center classification date
Jun 30, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records