Stryker Tornier Perform Humeral System, Reversed Insert, 45%, REF DWP2393, for shoulder arthroplasty — FDA record Z-3309-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Aug 29, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Stryker Tornier Perform Humeral System, Reversed Insert, 45%, REF DWP2393, for shoulder arthroplasty
Reason
Devices from one lot of Perform" Reversed Inserts may contain a locking ring that was assembled upside down which would prevent the humeral insert from securely seating in the stem or spacer.
Classification
Class II
Recalling firm
Tornier, Inc
Quantity
38 units
Agency-reported status
Ongoing
Source record identifier
Z-3309-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityGTIN 00846832084494, Batch Lot # 1495AY

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Oct 9, 2024
Recall initiation date
Aug 29, 2024
Center classification date
Sep 27, 2024