ADVIA Centaur aTPO Assay, Material Number 10630886 (100 Test) and 10630887 (500 Test) — FDA record Z-3304-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Aug 21, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
ADVIA Centaur aTPO Assay, Material Number 10630886 (100 Test) and 10630887 (500 Test)
Reason
Positive bias in patient results at and below the cut-off of 60 U/mL with the lots listed in the table below when compared to alternate lots on the Atellica IM analyzer.
Classification
Class II
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
5622 kits
Agency-reported status
Ongoing
Source record identifier
Z-3304-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: Material Number 10630886 (100 Test) - 00630414587837; Material Number 10630887 (500 Test) - 00630414587844; Lot Number: 26053337 26054337 26051337 26052337

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Oct 9, 2024
Recall initiation date
Aug 21, 2024
Center classification date
Sep 27, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.