LINK Universal Handle, with quick coupling, Stainless Steel, straight, Item Number 130-394/01 — FDA record Z-3245-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 11, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
LINK Universal Handle, with quick coupling, Stainless Steel, straight, Item Number 130-394/01
Reason
Some lots of the LINK Universal Handle, with quick coupling, Stainless Steel, straight may suffer an immediate material failure intraoperatively when the instrument is used for the first time. The instrument can break in the middle.
Classification
Class II
Recalling firm
Waldemar Link GmbH & Co. KG (Corp. Hq.)
Quantity
32 devices
Agency-reported status
Ongoing
Source record identifier
Z-3245-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 04026575215539; Lot Codes: C204156; C219063; C219064; C233033; C233034; C332135

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of FL, KS, MI, NJ, VA.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Oct 2, 2024
Recall initiation date
Sep 11, 2024
Center classification date
Sep 24, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.