Good Vibrations Ultrasonic Cleaner Solution, JorVet J0382G Ultrasonic Instrument Cleaning Solution, SpecClean Ultraso... — FDA record Z-3209-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Aug 29, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Good Vibrations Ultrasonic Cleaner Solution, JorVet J0382G Ultrasonic Instrument Cleaning Solution, SpecClean Ultrasonic Cleaner Solution, Models: GVUS128-1, GVUS128-1EA, GVUS550-1, GVUS24-1, GVUS24-1EA, JGVUS128-1, VGVUSNP128-1, VGVUSNP128-1EA
Reason
Contaminated equipment and packaging caused Klebsiella oxytoca, Enterobacter asburiae, Enterobacter bugandensis, pathogenic and opportunistic bacteria, to occur in ultrasonic detergent, which caused product bloating/leaking, and if used during medical/dental procedures could result in urinary tract, respiratory, bloodstream infections; workers handling contaminated product/instruments also at risk
Classification
Class II
Recalling firm
Microcare Medical
Quantity
1649
Agency-reported status
Ongoing
Source record identifier
Z-3209-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLot/Model: 2331982/GVUS128-1EA, GVUS128-1, GVUS550-1; 2332204/GVUS128-1EA, JGVUS128-1, GVUS128-1, VGVUSNP128-1EA, VGVUSNP128-1; 2332393/GVUS128-1EA, VGVUSNP128-1, GVUS24-1EA, GVUS24-1, GVUS128-1 2432755/JGVUS128-1 2432989/GVUS24-1EA, VGVUSNP128-1EA, GVUS24-1, VGVUSNP128-1, JGVUS128-1 Model/UDI-DI: GVUS128-1/+H816GVUS128H, GVUS24-1/+H816GVUSU24%, GVUS550-1/+H816GVUS550G, VGVUS128-1/+H816VGVUSNP128A, JGVUS128-1/723754044532

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution in the states of MN, GA, PA, WA, CO, NY, MD, CA, MI, NE, UT, OK, WI, NJ, IL, ME, CT, KS, TX, NC, TN, SC, MO and the countries of Canada.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Oct 2, 2024
Recall initiation date
Aug 29, 2024
Center classification date
Sep 23, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.