Ingenia Elition X- For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spect... — FDA record Z-3200-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 3, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Ingenia Elition X- For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 1) 782151; 2) 782119; 3) 781358; 4) 782107; 5) 782136
Reason
Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.
Classification
Class II
Recalling firm
Philips North America Llc
Quantity
33 units
Agency-reported status
Ongoing
Source record identifier
Z-3200-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityModel UDI-DI: 782151 N/A; 782119 (01)00884838104129(21); 781358 (01)00884838088115(21); 782107 (01)00884838098336(21); 782136 (01)00884838108608(21) Serial Numbers: 26500 26505 28502 28509 45393 45467 45599 62029 62041 62051 62052 62061 62066 62070 62071 62081 62082 62083 62091 62103 62111 62113 62114 62120 62126 62129 62147 62148 62186 62187 62188 66001 66002

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Iceland, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Maldives, Mauritius, Mexico, Morocco, Netherlands, Pakistan, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Switzerland, Thailand, T¿rkiye, United Kingdom, Utd.Arab.Emir., Uzbekistan, Vietnam.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Oct 2, 2024
Recall initiation date
Sep 3, 2024
Center classification date
Sep 20, 2024