Ingenia Ambition X - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spe... — FDA record Z-3198-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 3, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Ingenia Ambition X - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 782109
Reason
Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.
Classification
Class II
Recalling firm
Philips North America Llc
Quantity
63 units
Agency-reported status
Ongoing
Source record identifier
Z-3198-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: (01)00884838098350(21) Serial Numbers; 47558 47602 47606 47608 47609 47614 47615 47617 47621 47622 47625 47626 47628 47630 47634 47644 47646 47651 47654 47656 47657 47660 47665 47666 47667 47668 47671 47672 47674 47677 47678 47679 47680 47681 47684 47685 47687 47688 47694 47695 47697 47698 47700 47701 47702 47703 47705 47706 47708 47709 47710 47716 47733 47737 47738 47753 47755 47759 47785 47791 47800 47802 47804

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Iceland, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Maldives, Mauritius, Mexico, Morocco, Netherlands, Pakistan, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Switzerland, Thailand, T¿rkiye, United Kingdom, Utd.Arab.Emir., Uzbekistan, Vietnam.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Oct 2, 2024
Recall initiation date
Sep 3, 2024
Center classification date
Sep 20, 2024