Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Model Number C21510, RSM Pneumatic Tubing Kit item #2, REF C111... — FDA record Z-3180-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jul 10, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Model Number C21510, RSM Pneumatic Tubing Kit item #2, REF C11137; in vitro diagnostic instrument
Reason
Beckman Coulter is recalling the DxI 9000 Access Immunoassay Analyzer due to undersized Pneumatics Tubing in the reagent storage module causing air leaks, compressor failures or sample processing errors. Four complaints were filed for this event that led to Beckman Coulter finding undersized tubing in stock and initiated corrective actions. Use of the affected device may result in errors leading to delay of results.
Classification
Class II
Recalling firm
Beckman Coulter, Inc.
Quantity
170 units
Agency-reported status
Ongoing
Source record identifier
Z-3180-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI/DI 15099590230623, Serial Numbers: 300116, 300117, 300118, 300123, 300126, 300128, 300129, 300130, 300131, 300133, 300134, 300135, 300136, 300137, 300138, 300139, 300141, 300142, 300143, 300145, 300147, 300148, 300149, 300150, 300151, 300152, 300153, 300154, 300155, 300156, 300157, 300158, 300160, 300161, 300162, 300163, 300164, 300166, 300167, 300168, 300169, 300170, 300171, 300172, 300174, 300176, 300178, 300179, 300180, 300181, 300182, 300183, 300188, 300189, 300190, 300192, 300193, 300194, 300195, 300196, 300197, 300198, 300199, 300200, 300201, 300202, 300203, 300204, 300205, 300206, 300207, 300208, 300209, 300210, 300211, 300212, 300213, 300214, 300215, 300216, 300217, 300218, 300219, 300220, 300221, 300222, 300223, 300224, 300226, 300227, 300228, 300229, 300230, 300231, 300232, 300233, 300234, 300235, 300236, 300237, 300238, 300239, 300240, 300241, 300242, 300243, 300244, 300245, 300246, 300247, 300248, 300249, 300251, 300252, 300253, 300254, 300255, 300256, 300257, 300258, 300260, 300261, 300263, 300264, 300266, 300269, 300270, 300271, 300272, 300273, 300274, 300275, 300276, 300277, 300280, 300281, 300282, 300284, 300285, 300286, 300287, 300291, 300292, 300293, 300294, 300295, 300297, 300307, 300308, 300310, 300312, 300313, 300314, 300315, 300316, 300317, 300319, 300320, 300321, 300322, 300323, 300324, 300325, 300326, 300327, 300328, 300329, 300330, 300332, 300333

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Brazil, Croatia, Czech Republic, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, United Kingdom.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Sep 25, 2024
Recall initiation date
Jul 10, 2024
Center classification date
Sep 19, 2024