STREAMLINE BLOODLINE SET FOR DIALOG DR-Tubing to be used to transfer blood between a patient and a haemodialyser duri... — FDA record Z-3179-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Aug 15, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
STREAMLINE BLOODLINE SET FOR DIALOG DR-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2010M2096A
Reason
Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method
Classification
Class II
Recalling firm
B Braun Medical Inc
Quantity
11262112 units
Agency-reported status
Ongoing
Source record identifier
Z-3179-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI 04046955674992; All Unexpired Lot Numbers

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Sep 25, 2024
Recall initiation date
Aug 15, 2024
Center classification date
Sep 19, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.