Centricity PACS V2, V3, V4, V6, V7, software product is intended for the storage, reading, diagnostic review, analysi... — FDA record Z-3168-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jul 31, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Centricity PACS V2, V3, V4, V6, V7, software product is intended for the storage, reading, diagnostic review, analysis, annotation, distribution, printing, editing, and processing of digital images and data acquired from diagnostic imaging devices, All versions
Reason
Under certain conditions historical imaging data migrated from a legacy system may become incorrectly associated with a different patient. If this occurs, studies or images may be displayed under the wrong patient record, which could lead to misdiagnosis or a delay in patient care.
Classification
Class II
Recalling firm
GE Healthcare
Quantity
1443 systems
Agency-reported status
Ongoing
Source record identifier
Z-3168-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: V4 - 00840682124430, V6 - 00840682104807, V7 - 00840682145572; System IDs: CARTI02312611 CARTI22214087T 928757VPASW 928757VPBSW 602344PVSW2 602344REPSITEB MARICOPAC02308029 602344TIMSVM 441236VPSW 805395PVSW 805395PVTSW 310349PACSSWPV UCLAMC2259090 310349PACSSWBPV UCLAMC2259063 916781CPACSSRWT PALMDALEM01810792 PALMDALEM16023955 PALMDALEM20508297 909788HCAPACS1 916781CPACSCKS 916781CPACSBCKS GSHCACPSW GSH320037250 GSHCACTPSW RANCHO02255039 RANCHOPISU2 RANCHOPTESTISU 805370HCALR VMPACS69 LOSROBLES19666790 MCANPACSW CENPACSW 303584HCAPVSW 303584REPBSW 303584HCAPVSWTESTIMS 970476PVSW VVMC15111181 VVMC27255473 302674PACSSW 302674PACSBSW 302736SAEATSW JFKSW JFKR19980220B JFKSWOPT 727462PVSW 727462PVBSW MPMH2259089T 772461HCAPACSW 772461HCAPACSW1 772461HCAPACSWT 352333PACSSW NFRH19663978B 904202BMCDSW 904202BMCSSW 904202BMCDTSW MH826898334 407518COREPACS 407518CPREPB KRMCOREPACSSW KRMC19478073 KRMC25216192 615344HCASW PCMC420084442 WFLHIV01 WESTFLORI19661467 772335PACSSW 772335PVSW 850875PVSW2 904819PVSW 904819PV2SW 904819PVTSW 727519PVBSW 850325PACSCORE CRMC820084433 888663CPBSW 727519PVSW 888663CPTSW 888663CPB2SW 727519PVTSW VAMAF02345966 727519PVTSW-MOCK 407518SCPACSHC 850325HCASWOPT WFLHIVT01 PCMC42259036 352333PACSTSW WELLINGTON01810730 WELLINGTON19678279T WELLINGTON20036598 MJPACSW TRMPACSW TRMNC20021605 100115-8-Centricity PACS-21222171 TRMTESTPACS Doctors Ho22222417 MJPACDRSW 706651HCACORE 912466PVSW SGHS2259039 912276PVSW 912276PVTSW 706278PACSSWPV 706278PACSSWBPV 706278PACSSWTPV VSMQL01987331 404325PACSSWPV 406238PVSW 641422PVSW PROSPECT2259084 PROSPECT22128566 319272SCPSWULP 319272CPDRSW 319272CPDRBSW 319272SCPTSWULP 208COREPVSW EID19660790 208CORETIMS NORTHWEST25297966T 773880PVSW 795079PACSEESW1 795079PACSEESW2 312996PVSW2 JSHPACSSW 312996PVSW1 312996PVSWT1 312996PVSWT2 795079PACST JHSJHOCC02259030 773880PVBSW 773880PVTSW JSHPACSTSW 815285PACUV100 ADVPACSTSW1 847437PVSW 847437DCPVSW 847437SAPVTSW LaitekDataMigrationCPACSLabCPACS 847990PVSW S046PACSSWPV S801PACSSWPV S046PACSSWTPV ADVPACSSW ADVPACSBSW 219325PVSW1 219325PVBSW1 219325PVTSW1 765983PVSW 765983PVSW2 RHHCS02253964T 812232PACSPV 812232PACSTPV THRH20084364 FRNKHAPACSPV FRNKHAPACSTPV FRANKFORT19633100 RRMCPACSW RGMCPACSTW RRMC120080697 VTP01810617T VHT02334235P VTP19617532 985230PVSW NOMC2259032 CFO18227240 985230PVSWS 504903CPACSSW 504903CPACSTSW 504903CPACSIMS 781979PVSW 781979PVBSW 413447BCPVSW 413447HCPVSW 413447PVSW 413748PVSW 413748PVTSW 413748PVBSW 301698PVSW 301698PVTSW NORTHE02315652 NORTHE02370307T 517353PACSEESW 517353PACSEETSW 906387VPSW 248858PVASW 248858PVBSW 248858PVTSW 507288PVSW 507288PVSWT-TESTING OMCR19625746 651855PVSW 651855PVBSW 612831PVTSW 660885PVSW 660885PVTSW S011PACSSWPV S011PACSBSW S011PACSTSW S009PACSSWPV S009PACSSWBPV S009PACSSWTPV2 GPMCCOREPACS 601288PACSSWPV DBCSW 406345SCPACSTSW 704660PVSW 704660PVTSW 402484PACSSW AMIM02031832 609652PACSSW 609652PACSBSW 609652PACSTSW VHGPVSW VHGPVBSW VHGPVTSW UMDNJPVPSW UMDNJPVPBSW UMDNJPVPTSW 505609PVSW SJRMIMST20655929 VSMAI01812924 702733CPACSSW SHPACSW SUNPACSW VALLEYHEA14460287 SUNPACSREPB 20910879CPACST NORTHERNN20284710 NORTHERNN20838223 NORTHERNN20838225 STO01440342 BLH01442255TST BLH01451970 STO02157029 STO02157030T2 718206PVSW 718206PVTSW 2435-1-Centricity PACS-20608371 718206REPBSW 315798PVSW1 315798PTISU MSKPVSW NEWYORK01976546 MSKPVBSW MSKPVTSW2 MSKPVTSW3 ERM16295285 718920PVSW 718920PVTSW1 516763PVSW 516763PVBSW 516763PVTSW1 STO18806599 914357PVSW 914357PVBSW 914357PVTSW 315470CPACSASW 315470CPACSBSW 315470CPACSTSW 914681PVSW 914681PVTSW 718920PVTSW2 330543PACSSWPV CHPA2259061 CHPA2259088T 614234PVSW 614722PVSW 614722PVBSW SAH52259083 SAH52259110T 740348PACSSW 740348PACSBSW 740348PACSTSW 419383PVSW MUOO2259017 419383PVSWB STMARYS-TISU STMARYS-PISU S040PACSSWBPV S040PACSSWAPV S040PACSSWTPV 503257PACSSWPV UHS101383446 215913CPBSW 484526PVSW1 484526PVTSW 484526CPREPB 215710PVSW SMMC14898807 QUANTUMIM14640640A QUANTUMIM14776702B QUANTUMIM14776726T 570523PVSW EVANVTPACSIMS 724728SSASW 724728SSABSW 724728SSATSW2 412575CPACSSW 412623CPBSW 215453PACSSWPV GVH102398920T 401444PVBSW 401444PVSW 401444PVTSW AIMC01810707 AIMC19632995 864716PACSSWPV 864716PACSSWT 864716PACSSWBPV 803461PACSSWPV SCOA2259023 843347PVSW 843347PVTSW CNOS02186078 TEST PQM 423493PACSW 423493REPB WTN00104692T WTN2259070 WTN2259097 423837HCASWSU 423431PVSW 865977SSASW 865977SSTSW 806354CPACS 806354CPACSB 806354CPACST7 972906GSPACSTSW 361694VPSW DRISCOLLC02257225TIMS 214820EEPVSW1 SCW2259068 FORTDUNCA01810773 FORTDUNCA20854738T FORTDUNCA21337445 DSMCCOREPACSSW DSMCCOREPACSDRSW DSMC-22886533 972906GSPACSSW HCA-NORT02332691 MEDIC01812756 281274MSLPVSW 713441MHPVSWA 713704MHMSWA 713704MHMSWB 713704MHMSWTA 713704MHMSWTB 713741MHHPVSWB 832522MWHPVSW DHOL01810756 DHOL15291375T DHOL20631410 DHOL23253841 214820EEPVSW2 254724PVSW SCW2259095 361572SSASW CMDC2259015 801DASSW 801DSWSP1 801SOFTWARE 801SOFTWAREOPT SMH820071869 SMH8-26535851 SMH8-26535857 LGMSW LGHI20084372 STSP21736781 802775SSASW 802775SSATSW4X 206292PACSSWPV 206545EESW UWH20021873REPB 414465PVSW 414465PVTASW 414465PVTBSW 414465PVBSW 304487PVSW PCH12259035-TISU DZ1906PP02 DZ27??PP03 DZ2112CC01 HCIT1555339PACS HCIT4825134PACS HCIT4691855PACS HCIT5184728CPACS HCIT4859945PACS 0910553037 0910553037 ANZVICCAUL ANZVICSAND AH0104CS01 AH5045CS01 BH1038PC01 22303605 010120ST03 010120ST52 300013ST11 HCIT4691866CPACS HCIT4691855CPACS HCIT4759526PACS 4863935_CPACS HCIT7102800PACS HCIT4852374PACS-IMS HCIT5245250CPACS HCIT4860936PACS HCIT5317135CPACS HCIT4290686PACS HCIT5245275CPACS HCIT5471924PACS HCIT7104315PACS HCIT5245248CPACS HCIT5245248TESTCPACS HCIT7229698PACS 604520REPSITEA 604520TESTIMS 604586REPSITEB 709737PACS 709752IMSAsw 709752IMSBsw 709752GEPACS 709752DCSDAS02sw 709752DCSDAS04sw 709752TESTIMSsw 709752REPSERVERsw 709752DCSDAS01sw 709752DCSDAS03sw 709752DASNLA01 709752DASNLA02 709752DASNLA03 709752DASNLA04 709752DASNLA05 709752DASNLA06 709752DASNLB01 709752DASNLB02 709752DASNLB04 709752DASNLB05 709752DASNLB06 709752DASNLB07 709752DASNLB08 709752DASNLB09 709752DASNLB10 709752DASNLB12 709752DASNLB13 709752DASNLB14 709752DASNLB15 709737UTILITYsw 709752ITS2sw 709752ITS1sw 709752CORDMODsw 709651DASGA03 709292PACS 519426IMS 519426TESTIMS 519426GEPACS 709454PACS 905522GEPACS 905521GEPACS 905521IMSA 905522IMSB 905521TESTIMS 905575IMS 613732GEPACS 613732IMS 613732TESTIMS 416813REPSITEB 416813REPSITEA 416813TESTIMS 416813GEPACS 416326IMS1 416326IMS2 416326GEPACS 905372GEPACS 905372IMSA 905372IMSB 905372TESTIMS HCIT4602438PACS HCIT4920063PACS HCIT4096494PACS HCIT4601968PACS HCIT5026141PACS HCIT5183890CPACS 082499310155 082455150052 082455150052 082455160221 082455160221 082499180109 082499180102 082499180106 082455160255 082427190180 082445110005 082455190146 082499130790 082499130125 082499130790 082499020074 082499100188 082455040339 082407040684 082407040684 82455040334 082455040339 082455040147 082455040352 082445040028 082455040317 SiteID-Centricity PACS-20608297 082407040684 082499110183 82499220115 082499100382 082455150074 082455150068 082455180040 082499180080 082499010124 082499060054 082499100386 082499100386 na 082499100461 082455100286 082499250135 082455070037 082455120548 082455120548 082455100352 082455100179 082455100179 082455160241 082433120225 082499120374 082455120507 082499250135 082455120545 082455120552 082455120510 082499100188 082455130065 082455130335 82455290014 082499130780 082455130366 082455130394 082455290015 082499180085 082445030013 082455160035 082499130792 082455120226 082455080025 082455060004 082455050079 082455050076 082455050069 tbd 082455130415 082427150109 082455120544 13767625-Centricity PACS-21273767 082499130816 082499130826 082455080026 082407120583 082455100357 082499100415 082499230063 SH EASTERN HEP24523158 HCIT50000107PACS HCIT4946501PACS DoNotUse-DK9154CA01 DK9066IMS1 DK9066TESTPAC1 DK9066IMS2 DK9066TESTPAC2 HCIT4277575PACS HCIT4692615PACS HCIT4796839PACS EG1526PP02 EG1017PP02 HCIT5162696CPACS ZA2879PP01 M4477025 A4167665 22106615 A4498636 HCL_24968096 M61421909 M100323902 M2030921 M2753440 M4171826 M4476437 X59655539 Not Available M4206540 M5172536 M5172537 M5172568 P4172826 P4172827 MQ1234502 M386578303 A27411 M4144112 M4045536 M4046544 M4047061 M40480163 M4188439 M54807112 NC1028 M4491536 PF1000PC02 M143153109 M4143924 M4143984 M220995201 02398902 M4825009 / M4825010 / M4825011 M4204530 H4858984 D4858905 D4858906 D4858918 M838835602 M4209925 M4209926 M4209927 Utuora HNO M4505229 HC2641CS01 Not Available 744459PA02 HC0220CS01 518086CS01 HC2720CS02 HC4017CS01 UC2847CS01 UESS01CS01 126292CS01 UJJF01CS01 416005CS01 HC4912CS01 530366PC03 HC0468CS02 556162CS01 UHAM02CS02 HC0727CS01 tobecreated UHAN01CS01 UHAN01CS01P HHEI01CS01 HC7048CS02 410014SV02 HC1992CS01 HC2554CS01 tobecreated 624751CS01 HC0014CS01 Not Available HC1467PA01 560496CS01 145774CS01 HCNE01CS01 HC0094CS01 553160CS01 434487CS01 HCSC01CS01 UC4372SV02 HC0863CS01 Not Available HC1282CS01 424881PA01 0828550074 0828550085 0828550070 828550079 0828550105 0828550072 0828550068 0828550068 0828550068 828550052IB01 828550064 828550066 828550052IB01 828550095 0828550073 0828550111 0828550071 0828550071 0828550071 0828550086 0828550086 828550057 083055283270418 MAX-DC-PACS 083055003435816 083055247345516 083055820666618 MCC-CPACS SCTIMST-IN-CPACS RCC - CPCS 7.0.0.4.8 083055201766451 SIRHNHOSPITALTRUST-CPACS Wockhardt-CPACS7.0.0.4.8 INHS-ASWINI-CPACS 083055211206517 ACTREC26874950 AIIMS-NM-PACS JIPMER-CPACS Not Available 083055803467916 083055801005510 SRK-CPACS 083055808240917 NARUVI -CPACS 083055840003710 CARE-DC-CPACS CHL-CPACS TMCV_CPACS 083055604732915_CPACS RSMITRAKELUARGABINTARO-CPACS RSUDBERKAHP-CPACS RSMKG-SERPONG01 BETHSAIDA-CPACS RSUDMukomuko-CPACS RSUDSOREANG-CPACS RSUDMadaniPalu-CPACS RSMKKEMAYORAN-CPACS RSUDKOTAJOGJA-CPACS MayapadaBandung-CPACS TEASIAHEALTHCARE-CPACS RSHARAPANKITA-CPACS RSKELUARGAKALIDERES-CPACS RSUDTarakan-CPACS PTMKMas02 PTSentosa1 RSUPFATMAWATI-CPACS RSPOLRIB-CPACS RSMK-KelapaGading01 SULIANTI001 PTUROGEN01 SiteID-Centricity PACS-20535377 RSUPTVertikalKupang-CPACS RSUDHMB-CPACS SANTOSA-CPACS HASANSADIKIN-EA RSMK-Cibubur01 RSUDBEKASI-CPACS MKCikarang-CPACS MKBekasiBarat-CPACS MKBEKASITIMUR_CPACS RSUDCIAWI-CPACS RSMITRAKD-CPACS SYAMSUDIN001 RSUDPandanArang-CPACS RSUDDRMAP-CPACS RSUDKRMTW-CPACS RSJDSurakarta-CPACS RSUDSlawi-CPACS RSDDRSJ-CPACS RSUDPasirian-CPACS RSUDSidoarjo-CPACS RSMITRAKW-CPACS RSALDr.Ramelan-CPACS RSHajiSurabaya-CPACS RSPHCSurabaya-CPACS RSUDTuban-CPACS RSUDKabupatenB-CPACS RSUDTarakanKaltara-CPACS RSUDLeimena-CPACS RSReginaMaris-CPACS RSUDKlungkung-CPACS RSPBINTARO2 MITRAKELUARGATEGAL-CPACS RSUDDRHKB-CPACS PTUNISIAMEDIKAFARMA_CPACS RSPJATINNEGARA1 MitraKeluargaSlawi-CPACS RSMardiRahayu-CPACS RSUDBHAKTIDHARMAHUSADA-CPACS RSUPSOERADJI-CPACS Dr.MOHAMMADZYN_CPACS MitraKeluargaDeltamas-CPACS RSMKSS-CPACS RSMITRAKK-CPACS RSUPDr.Wahidin-CPACS RSUDPadangan-CPACS RSPSURABAYA1 PTSEJANU01 MitraKeluargaPamulang-CPACS RSMITRA-CPACS RSUDProvinsiNTB-CPACS RSUDKOTADUMAI-CPACS RSUDKKM-CPACS RSUDTelukKuantan-CPACS RSOtak-CPACS RSUPM.DJAMIL-CPACS RSUDPalembang_CPACS HoesinPALEMBANG-CPACS RSUDSIJUNJUNG-CPACS MitraKeluargaJatiasih-CPACS RSBinaHusadaCibinong-CPACS RSUDPANDEGA-CPACS IQ1906PC03 IQ1727PC03 81100DAS05 IL1018IT01 IL1059IT01 Not Available IL1484IT01 IL1469IT08 IL1098IT10 01423795 01452223 A51043144 A5802803 C5368613 A5368605 A5801809 B5368603 A5630621 01452224 01979741 A5353013 01452218 A5660044 A6085807 01452220 C5603910 A5105147 A670060230 A5156018 A5119431 A5719805 26648960 B5105141 JL1748 YP5025-Centricity PACS-02170316 JA0842 JDS021 JL1715 JAS007 JDS028 JDS029 JDS030 JL1726 JAS023 JDS084 JDS085 JDS139 JL1744 JL1717 JAS009 JDS035 JDS036 JDS037 JDS043 JL1754 JA0832 JD1164 JD1165 JD1166 YP4012-080218-Centricity PACS-20579603 JL1743 JAS041 JL1751 JD1275 YP5018-Centricity PACS-02013203 JL1755 JAS039 JL1739 JAS035 JAS022 JDS082 JDS083 YP5040-Centricity PACS-19737087 JL1741 JAS036 JL1393 JD1015 JD1016 JDS133 YP5047-Centricity PACS-20617590 YP5044-Centricity PACS-20484083 JAS027 Y...

The agency lot list is longer than this excerpt. Read the complete list at the official source.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Sep 23, 2026
Recall initiation date
Jul 31, 2026
Center classification date
Sep 15, 2026