Ivenix Infusion System (IIS), LVP Software; LVP-SW-0004, Software Version 5.9.1 and prior. Infusion pump software for... — FDA record Z-3153-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Aug 30, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Ivenix Infusion System (IIS), LVP Software; LVP-SW-0004, Software Version 5.9.1 and prior. Infusion pump software for infusion management system.
Reason
The software has anomalies that have the potential to cause alarms, nonfunctioning pump, or unresponsive screen while continuing therapy. These could cause serious patient harm or death.
Classification
Class I
Recalling firm
Fresenius Kabi USA, LLC
Quantity
17 units
Agency-reported status
Ongoing
Source record identifier
Z-3153-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 00811505030122; Software Version 5.9.1 and prior

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Oct 2, 2024
Recall initiation date
Aug 30, 2024
Center classification date
Sep 26, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records