Seer Home System is an ambulatory electroencephalograph (EEG) system intended for at-home use for extended EEG and vi... — FDA record Z-3143-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Aug 12, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Seer Home System is an ambulatory electroencephalograph (EEG) system intended for at-home use for extended EEG and video EEG (vEEG) recordings and analysis. The Seer Home System includes the Sight Monitoring Hub, the Sense wearable, Monitoring Portal and associated accessory components (Disposable EEG Wires, WaterTabs, WaterTabs Application Kit, and Take-home Kit).
Reason
During CAPA investigations of sight production inconsistencies, it was highlighted that previous Electromagnetic Compatibility (EMC) assessments related to RF radiated emissions may not be reasonably equivalent to the current product due to design changes in the system. A new CAPA was opened to investigate the radiated emissions compliance issues. - Seer Home Systems distributed and retained in the field were not compliant with Electromagnetic Compatibility (EMC) testing according to IEC 60601-1-2 - Seer Home Systems, including accessories, were manufactured and distributed without application of proper design change and manufacturing controls - Seer Home Systems were distributed with labeling deficiencies
Classification
Class II
Recalling firm
SEER MEDICAL PTY LTD
Quantity
3,948 units
Agency-reported status
Ongoing
Source record identifier
Z-3143-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicability1. Product Number: SM-03-8001-INT; UDI-DI: 09355751001037; Serial Numbers: 000089, 000090, 000091, 000092, 000095, 000182, 000201, 000207, 000208, 000211, 000225, 000245, 000258, 000259, 000285, 000286, 000289, 000290, 000291, 000292, 000296, 000297, 000298, 000299, 000301, 000305, 000307, 000308, 000309, 000310, 000312, 000313, 000316, 000321, 000322, 000323, 000094, 000183, 00028, 000302, 000303, 000306, 000311. 2. Product Number: SM-01-8001; UDI-DI: 09355751001020; Serial Numbers: 000074, 000084, 000113, 000118, 000119, 000187, 000188, 000191, 000223, 000224, 000225, 000226, 000228, 000229, 000230, 000232, 000239, 000241, 000242, 000243, 000244, 000245, 000247, 000250, 000251, 000259, 000262, 000263, 000277, 000279, 000280, 000282, 000283, 000286, 000287, 000288, 000289, 000290, 000291, 000292, 000294, 000297, 000299, 000300, 000301, 000302, 000303, 000304, 000305, 000306, 000307, 000308, 000309, 000310, 000312, 000313, 000315, 000317, 000318, 000319, 000320, 000321, 000322, 000323, 000324, 000327, 000328, 000329, 000333, 000335, 000336, 000337, 000339, 000340, 000341. 3. Product Number: Monitoring Portal; UDI-DI: 09355751001136; Serial Numbers: DMPD80RCMDFX, DMPFRRE7Q1GJ, DMPFT3PDQ1KQ, DMPFT43HQ1GG, DMPFT46JQ1KQ, DMPFT4CYQ1KQ, DMPFT4X1Q1KQ, DMPFV0HGQ1KQ, DMPFV178Q1KQ, DMPFV19SQ1KQ, DMPFV1AFQ1KQ, DMPFV1AXQ1KQ, DMPFV1DBQ1KQ, DMPFV1KQQ1KQ, DMPFV1L2Q1KQ, DMPFV1LNQ1KQ, DMPFV1PRQ1KQ, DMPFV1Q8Q1KQ, DMPFV1R2Q1KQ, DMPFV1R9Q1KQ, DMPFV1SRQ1KQ, DMPFV1TEQ1KQ, DMPFV1WYQ1KQ, DMPFV1YGQ1KQ, DMPFV1ZBQ1KQ, DMPFV1ZMQ1KQ, DMPFV20DQ1KQ, DMPFV20PQ1KQ, DMPFV219Q1KQ, DMPFV21GQ1KQ, F9FF60LYQ1KQ, F9FF70Z5Q1KQ, F9FF71PAQ1KQ, F9FFQ413Q1GJ, F9FFQC4WQ1GG, F9FFQF9VQ1GG. 4. Product Number: SM-01-9005; UDI-DI: 09355751001112; Serial Number: 006240129. 5. Product Number: SM-07-8001; UDI-DI: 09355751001129; Serial Numbers: 007230809, 007230810, 007230823, 007230825, SM-07-9001_230829, 007231004, 007231005, 007231012, 007231023, 007231025, 007231102, 007231108, 007231109, 007231115, 007231120, 007231124, 00720240129, 00720240205, 00720240207, 00720240314, 007240318, 007240319, 007240327, 007240325, 007240326, 007240402, 007240403, 007240507, 007240515, 007240516, 007240522, 007240527, 007240531, 007240612, 007240613, 007240624, 007240626, 007240627, 007240628, 007240702, 007240703. 6. Product Number: SM-01-9003-US; UDI-DI: 09355751001099; Serial Numbers: 230525, 004230609, 004230913, 004231121, 004240306, 004240403, 004240424, 004240502, 004240509, 004240531, 004240603, 004240626, 004240705. 7. Product Number: SM-01-9008; UDI-DI: 09355751001105; Serial Numbers: 00223058, 00223059, 00223066, 00224008, 00224011, 00224013, 00224017, 00224018, 00224021.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution in the states of MN, NY, PA and the countries of Australia.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Sep 25, 2024
Recall initiation date
Aug 12, 2024
Center classification date
Sep 13, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.