Recall archive / FDA device enforcement
Baxter EXACTAMIX Inlet, Vented, High Volume Inlet, REF H938174, for use in a pharmaceutical compounding device — FDA record Z-3083-2024
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Aug 20, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Baxter EXACTAMIX Inlet, Vented, High Volume Inlet, REF H938174, for use in a pharmaceutical compounding device
- Reason
- Baxter Healthcare Corporation has received increased customer reports of particulate matter in the Automated Compounding Device Inlets (disposable inlet), Product Codes H938173, H938174, H938175, H938176, used with the ExactaMix and ExactaMix Pro compounders. Particulate matter has been observed within the inlet primary packaging inlet components, including within the sterile fluid path tubing, before use.
- Classification
- Class I
- Recalling firm
- Baxter Healthcare Corporation
- Quantity
- 1,548,782 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-3083-2024
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | UDI/DI 00085412475790, Lot Numbers: 803799 and lower |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide distribution.
No US distribution state could be established from this record. Check the official notice.
Dates in the source
- Report date
- Sep 25, 2024
- Recall initiation date
- Aug 20, 2024
- Center classification date
- Sep 19, 2024
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.