PORTRAIT CORE SERVICES SOFTWARE V1.0.5, Ref# 5697353 — FDA record Z-3038-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jul 20, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
PORTRAIT CORE SERVICES SOFTWARE V1.0.5, Ref# 5697353
Reason
Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area.
Classification
Class II
Recalling firm
GE Medical Systems Information Technologies Inc
Quantity
15 units
Agency-reported status
Ongoing
Source record identifier
Z-3038-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityGTIN 00195278364838, Serial Numbers: SUC23080050TA, SUD22470001TA, SUD22420001TA, SUD22470010TA, SUD23390001TA, SUD23390003TA, SUD23390004TA, SUD23390009TA, SUD23390010TA, SUD22470009TA, SUD22470003TA, SUD22470004TA, SUD22470005TA, SUD23390005TA, SUD22280002TA.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US: MI, NY and Australia, Czech Republic, Finland, Germany, Italy, Netherlands, Qatar, Saudi Arabia, Switzerland, United Arab Emirates, United Kingdom.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Sep 2, 2026
Recall initiation date
Jul 20, 2026
Center classification date
Aug 27, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records