25-LHP-828, HeNe Laser System, 35 mW — FDA record Z-2975-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Mar 1, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
25-LHP-828, HeNe Laser System, 35 mW
Reason
These laser products fail to comply with 21 CFR 1040.10(f)(4) Key control which requires, Each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. , because while the laser is switched on, the key can be removed from the key switch. The correction is replacement with a key switch that captures the key when in the on position.
Classification
Class II
Recalling firm
PACIFIC LASERTECH, LLC
Quantity
4
Agency-reported status
Ongoing
Source record identifier
Z-2975-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicability25-LHP-828

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide Distribution

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Sep 18, 2024
Recall initiation date
Mar 1, 2024
Center classification date
Sep 10, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.