Infusomat Space Infusion System/ Large Volume Pump, Product Code 8713050U — FDA record Z-2972-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Aug 7, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Infusomat Space Infusion System/ Large Volume Pump, Product Code 8713050U
Reason
Extension of previous recall RES 92978. Pumps have the potential for upstream occlusion sensor to cease proper function due to buildup of electrostatic charge during operation triggering false downstream and upstream occlusion alarms. Additional devices have been affected.
Classification
Class I
Recalling firm
B Braun Medical Inc
Quantity
14
Agency-reported status
Ongoing
Source record identifier
Z-2972-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI (GUDID) - 04046963716752 Serial Numbers 52226, 107529, 123259, 123815, 123843, 139003, 147678, 147721, 147725, 239018, 239165, 239231, 239279, 339337

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US distribution to AL, CA, GA, IL, MI, and TX.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Sep 18, 2024
Recall initiation date
Aug 7, 2024
Center classification date
Sep 6, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records