Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard I... — FDA record Z-2911-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jun 29, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0630-140-2, NG-0640-140-2, NG-0730-140-2, NG-0740-140-2, NG-NV-6-30, NG-NV-6-40, NG-NV-7-30, NG-NV-7-40
Reason
Carotid stent and post-dilation balloon system with integrated embolic protection system has an integrated filter, that due to a wire-to-filter-bond issue, may fail to open following sheath retraction; and if this condition were to occur during a procedure, the result may be loss of intended integrated embolic protection.
Classification
Class II
Recalling firm
Contego Medical, Inc
Quantity
5069
Agency-reported status
Ongoing
Source record identifier
Z-2911-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityREF/UDI-DI/Lot: NG-0630-140-2/00851616007583/Z2460578D,Z2460740B, Z2460872C, Z2460909F, Z2460948D, Z2460957Q; NG-0640-140-2/00851616007590/Z2460578E, Z2460647S, Z2460706A, Z2460746C, Z2460774A, Z2460865C, Z2460872B, Z2460901B, Z2460909D, Z2460909K, Z2460945H, Z2460948I, Z2460963N; NG-0730-140-2/00851616007606/Z2460578F, Z2460662A, Z2460740A , Z2460774F, Z2460850A, Z2460900B, Z2460945F, Z2460957R; NG-0740-140-2/00851616007613/Z2460578G, Z2460635C, Z2460656A, Z2460707D, Z2460733A, Z2460766A, Z2460774B, Z2460780A , Z2460865A, Z2460871A, Z2460872D, Z2460900D, Z2460909B, Z2460909I, Z2460948F, Z2460963T; NG-NV-6-30/00851616007644/Z2460776A, Z2460893A, Z2460909G, Z2460948E; NG-NV-6-40/00851616007651/Z2460776B, Z2460865D, Z2460865E, Z2460901C, Z2460909E, Z2460909J, Z2460957P, Z2460963K, Z2460963P; NG-NV-7-30/00851616007637/Z2460776C, Z2460789H, Z2460850B, Z2460900C, Z2460916A, Z2460948H, Z2460998A; NG-NV-7-40/00851616007620/Z2460776D, Z2460878A, Z2460872A, Z2460872E, Z2460901A, Z2460909C, Z2460909H, Z2460945G, Z2460948G, Z2460966B, Z2460963J;

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of IL, FL, NC, WI, KY, AL, VA, DE, NJ, PA, IN, MI, CA.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Aug 12, 2026
Recall initiation date
Jun 29, 2026
Center classification date
Aug 6, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records