Recall archive / FDA device enforcement
Atellica CH Urine Albumin REF: 11537225, in vitro diagnostic — FDA record Z-2890-2026
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Jul 6, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Atellica CH Urine Albumin REF: 11537225, in vitro diagnostic
- Reason
- Processing of urine albumin assay on the CI analyzers may lead to the following issues: Falsely depressed Atellica CH Urea Nitrogen (UN_c) results; Falsely depressed Atellica CH Enzymatic Hemoglobin A1c (A1c_E) results; Increased rate of reaction cuvette QC failures and Delayed results due to QC failures.
- Classification
- Class II
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 7018 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-2890-2026
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | All Lots within expiry/UDI: 00630414611099 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide distribution - US Nationwide including Puerto Rico and the countries of Australia, Austria, Bahamas, Belgium, Brazil, Canada, Chile, Curacao, St. Eus, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Ireland, Israel,¿Italy,¿Latvia, Malaysia, Netherlands, New Zealand, Norway, Portugal, Republic of Korea, Romania, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, United¿Kingdom.¿
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Aug 12, 2026
- Recall initiation date
- Jul 6, 2026
- Center classification date
- Aug 5, 2026