INFINION CX LEAD KIT, 50CM, Model: SC-2317-50, Material: M365SC2317500; INFINION CX LEAD KIT, 70CM, Model: SC-2317-70... — FDA record Z-2879-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jun 17, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
INFINION CX LEAD KIT, 50CM, Model: SC-2317-50, Material: M365SC2317500; INFINION CX LEAD KIT, 70CM, Model: SC-2317-70, Material: M365SC2317700. part of the Boston Scientific Spinal Cord Stimulator (SCS) Systems.
Reason
Spinal cord stimulation system lead body may be exposed to stress at the anchor site. This may result in high impedance measurements or lead fractures, which may lead to inadequate stimulation, additional intervention, including lead explant and replacement.
Classification
Class I
Recalling firm
Boston Scientific Neuromodulation Corporation
Quantity
1079
Agency-reported status
Ongoing
Source record identifier
Z-2879-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityModel/UDI-DI: SC-2317-50/08714729861614, SC-2317-70/08714729861638. All unused products.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US: NY. OUS: Canada, Portugal, Spain, France, Netherlands, Great Britain, Italy, Israel, Serbia, Greece, Germany, United Arab Emirates, Ireland, Hungary, Czech Republic, Brazil, Colombia, Argentina

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Aug 12, 2026
Recall initiation date
Jun 17, 2026
Center classification date
Aug 5, 2026