Brand Name: OXOS Medical, Inc Product Name: MC2 Portable X-Ray System Model/Catalog Number: MC2, MC2.1 Software Versi... — FDA record Z-2857-2026
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Jun 11, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Brand Name: OXOS Medical, Inc Product Name: MC2 Portable X-Ray System Model/Catalog Number: MC2, MC2.1 Software Version: 5.7.0 Product Description: The MC2 Portable X-Ray System ( MC2 System or MC2 ) is a portable and handheld x-ray system designed to aid clinicians with point of care visualization through diagnostic x-rays of extremities and shoulders. The device allows a clinician to select desired technique factors best suited for their patient anatomy. The MC2 consists of two major system components: the emitter and the cassette. The MC2 emitter and cassette are battery-powered and can be charged with either a wired or wireless charger. The system is intended to interface wirelessly to an external tablet or display when used with the OXOS Device App. The MC2 utilizes an infrared Tracking System and Viewfinder that allows the operator to position the emitter within a safe source-to-image distance (SID) and allows the operator to visualize the anatomy and x-ray field projection prior to initiating an x-ray acquisition. The MC2 also utilizes a LIDAR system to ensure patient safety by maintaining a safe source-to-skin distance. Component: The MC2 Portable X-Ray System is not a component.
- Reason
- Due to a defective circuit board that overheats
- Classification
- Class II
- Recalling firm
- Oxos Medical Inc
- Quantity
- 77
- Agency-reported status
- Ongoing
- Source record identifier
- Z-2857-2026
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | Lot Code: MC2; UDI 00860008892782; Serial Numbers 1523, 1527, 1488, 1528, 1529, 1544 MC2.1; UDI 00850079606043; Serial Numbers: 1461, 1462, 1547 Full UDI for affected devices: (01)00860008892782(11)260107(21)1523 (01)00860008892782(11)260129(21)1527 (01)00860008892782(11)260109(21)1488 (01)00860008892782(11)260123(21)1528 (01)00860008892782(11)260130(21)1529 (01)00860008892782(11)260227(21)1544 (01)00850079606043(11)260407(21)1461 (01)00850079606043(11)260407(21)1462 (01)00850079606043(11)260408(21)1547 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US Nationwide distribution in the states of FL, GA, IN, SC, TX, VA.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Aug 12, 2026
- Recall initiation date
- Jun 11, 2026
- Center classification date
- Aug 3, 2026
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.