- Product
- REMISOL Advance with Software Version 2.5 UDI-DI code: 15099590802073 The Normand Info REMISOL Advance Data Manager collects and manages data and workflow for 128 connections
- Reason
- Due to software issues where does not handle combo assay results. The software may replace the final combo test result with one of the sub-results resulting in incorrect reporting of test outcomes.
- Classification
- Class II
- Recalling firm
- Beckman Coulter Biomedical GmbH
- Quantity
- 20 systems
- Agency-reported status
- Ongoing
- Source record identifier
- Z-2834-2026
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | Software Version 2.5 UDI-DI code: 15099590802073 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
U.S.: FL, GA, IL, MI, and PA O.U.S.: Belgium, France, Hungary, Poland, Taiwan (Province of China), Tunisia, United Arab Emirates, and United Kingdom of Great Britain and Northern Ireland.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
Dates in the source
- Report date
- Aug 12, 2026
- Recall initiation date
- Apr 29, 2026
- Center classification date
- Jul 31, 2026
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.