What the record says

Product
EOPA 3D Arterial Cannula 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 78322
Reason
Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.
Classification
Class II
Recalling firm
Medtronic Perfusion Systems
Quantity
1,424 units
Agency-reported status
Ongoing
Source record identifier
Z-2811-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 20199150023251, Lot Numbers: 0232709184 0232745036 0232798416 0232862784 0232862787; UDI-DI: 00199150023257, Lot Numbers: 0232862787

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Aug 5, 2026
Recall initiation date
Jun 24, 2026
Center classification date
Jul 27, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.