EOPA Elongated One-Piece Arterial Cannula BLUNT 22FR, intended for use with cardiopulmonary bypass as an arterial ret... — FDA record Z-2807-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jun 24, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
EOPA Elongated One-Piece Arterial Cannula BLUNT 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77522
Reason
Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.
Classification
Class II
Recalling firm
Medtronic Perfusion Systems
Quantity
11,720 units
Agency-reported status
Ongoing
Source record identifier
Z-2807-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 00199150023288, Lot Numbers: 0233286436; UDI-DI: 20199150023282, Lot Numbers: 0232709428 0232744924 0232745061 0232745092 0232798556 0232862763 0232862778 0232862794 0232862798 0232903216 0232903291 0232903383 0232990983 0233286318 0233286414 0233286436 0233427120 0233528730 0233553623 0233556386 0233585113 0233585114 0232670802

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Aug 5, 2026
Recall initiation date
Jun 24, 2026
Center classification date
Jul 27, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records