SafeOp Patient Interface, REF: AIX11240, part of SafeOp 3: Neural Informatix System — FDA record Z-2793-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 13, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
SafeOp Patient Interface, REF: AIX11240, part of SafeOp 3: Neural Informatix System
Reason
Intraoperative neuromonitoring device, due to a software issue, may experience loss or interruption of triggered electromyography (tEMG) functionality during use, displaying as system faults, loss of communication between the Patient Interface and Tablet, and prompts requiring system reboot; may result in increased operative time and surgical plan change.
Classification
Class II
Recalling firm
Alphatec Spine, Inc.
Quantity
3
Agency-reported status
Ongoing
Source record identifier
Z-2793-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 00190376517506. System Set ID/Patient Interface Serial Number: 2007716/ DG106225, 2005498/ DG106220, 2005490/ DG106230

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of TX, MO, IN.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jul 29, 2026
Recall initiation date
May 13, 2026
Center classification date
Jul 23, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records