FARADRIVE STEERABLE SHEATH CLEAR OUS, used to facilitate access to the vasculature and into the chambers of the heart... — FDA record Z-2757-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jun 11, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
FARADRIVE STEERABLE SHEATH CLEAR OUS, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PFCE21M402, IFU Reference Number 51659727-01
Reason
Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
Classification
Class II
Recalling firm
Boston Scientific Corporation
Quantity
193,000 units
Agency-reported status
Ongoing
Source record identifier
Z-2757-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 00191506042516; Lots: CL12917, CL13004, CL13017, CL13046, CL13060, CL13080, CL13103, CL13122, CL13123, CL13142, CL13160, CL13196, CL13212, CL13219, CL13231, CL13260, CL13291, CL13340, CL13375, CL13401, CL13422, CL13468, CL13519, CL13573, CL13578, CL13614, CL13646, CL13670, CL13671, CL13708, CL13719, CL13722, CL13732, CL13760, CL13801, CL13802, CL13820, CL13833, CL13856, CL13910, CL13911, CL13925, CL13940, CL13971, CL13972, CL13988, CL14025, CL14039, CL14050, CL14051, CL14076, CL14077, CL14119, CL14120, CL14135, CL14137, CL14171, CL14187, CL14226, CL14249, CL14286, CL14287, CL14345, CL14346, CL14375, CL14376, CL14396, CL14397, CL14439, CL14488, CL14500, CL14501, CL14518, CL14519, CL14581, CL14582, CL14633, CL14635, CL14670, CL14697, CL14711, CL14749, CL14750, CL14769, CL14770, CL14813, CL14814, CL14854, CL14855, CL14870, CL14871, CL14928, CL14929, CL14933, CL14934, CL14935, CL14968, CL14969, CL15005, CL15009, CL15067, CL15107, CL15108, CL15109, CL15128, CL15129, CL15191, CL15192, CL15193, CL15239, CL15240, CL15272, CL15318, CL15358, CL15397, CL15427, CL15474, CL15490, CL15539, CL15568, CL15569, CL15597, CL15598, CL15637, CL15638, CL15640, CL15754, CL15755, CL15775, CL15776, CL15808

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Global Distribution. US Nationwide.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jul 29, 2026
Recall initiation date
Jun 11, 2026
Center classification date
Jul 17, 2026