Recall archive / FDA device enforcement
Orthofix MINIRAIL, ARTICULATED, FIXATOR, VERTICAL AXIS REF: M122, Unilateral orthopedic/trauma external fixation device — FDA record Z-2732-2026
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Jun 10, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Orthofix MINIRAIL, ARTICULATED, FIXATOR, VERTICAL AXIS REF: M122, Unilateral orthopedic/trauma external fixation device
- Reason
- Unilateral external fixation devices may be incorrectly labeled. Minirail M101 was labeled as M103, Minirail M103 was labeled as M101, and Minirail M122 was labeled as M111
- Classification
- Class II
- Recalling firm
- Orthofix U.S. LLC
- Quantity
- 62 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-2732-2026
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | All Lots; GTIN# 18032568866711 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US: AZ CA CO CT FL GA ID IL IN LA MD MI MO MS MT NC NY OH OR SC SD TX UT VA WA WI OUS: None
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
Dates in the source
- Report date
- Jul 22, 2026
- Recall initiation date
- Jun 10, 2026
- Center classification date
- Jul 14, 2026
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.