Recall archive / FDA device enforcement
Stryker iBur 4.0mm Diamond Round, Distal Bend- intended to cut bone in the following manner: drilling, reaming, decor... — FDA record Z-2726-2024
Reported completed
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Jul 24, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Stryker iBur 4.0mm Diamond Round, Distal Bend- intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following medical applications: Neuro; Spine; Ear, Nose, and Throat (ENT)/ Otorhinolaryngology; and Endoscopic applications Ref: 8431-012-040D
- Reason
- Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.
- Classification
- Class II
- Recalling firm
- Stryker Corporation
- Quantity
- 1706 units
- Agency-reported status
- Completed
- Source record identifier
- Z-2726-2024
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | GTIN: 07613327501278 Lot Numbers: 22334017 22339017 23004017 23016017 23085017 23115017 23195017 23299017 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Canada, France, Hong Kong, India, Italy, Japan, Japan, Netherlands, South Korea, Spain, Sweden, Taiwan, United Kingdom.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Sep 4, 2024
- Recall initiation date
- Jul 24, 2024
- Center classification date
- Aug 26, 2024
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.