Brand Name: Brachytherapy Applicator Product Name: Advanced Gynecological Applicator Product Description: The intende... — FDA record Z-2711-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 3, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Brand Name: Brachytherapy Applicator Product Name: Advanced Gynecological Applicator Product Description: The intended use of the Advanced Gynecological Applicator is intracavitary brachytherapy for cancer treatment of the cervix and endometrium. Optional needles can be placed for interstitial brachytherapy.
Reason
Due to safety concerns regarding the detachment of the tip of the intrauterine tube (IU tube) that may remain within a patient.
Classification
Class II
Recalling firm
Elekta, Inc.
Quantity
586
Agency-reported status
Ongoing
Source record identifier
Z-2711-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLot Code: 08717213055001, 08717213055018, 08717213055025, 08717213055049, 08717213055056, 08717213055063, 08717213055087, 08717213055094, 08717213055100, 08717213055117, 08717213055131

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - U.S. Nationwide distribution including in the states of AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TX, UT, VA, WI, and WV. The countries of Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Botswana, Brazil, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Georgia, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iran (Islam Republic of), Iraq, Ireland, Italy, Japan, Kazakhstan, Kenya, Korea (Republic of), Kosovo, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Rwanda, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Swaziland, Sweden, Switzerland, Taiwan, Thailand, Togo, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, Viet Nam, and Zambia.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jul 15, 2026
Recall initiation date
Apr 3, 2026
Center classification date
Jul 9, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.