DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in sampl... — FDA record Z-2691-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jul 10, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only.
Reason
Due to a software issues, after the instrument processes 250 racks cumulatively, any subsequent racks with samples requiring rerun/reflex will be held in the Sample Handler's Buffer area and and error code "9000" will be reported. This issue may cause a delay of results.
Classification
Class II
Recalling firm
Beckman Coulter Mishima K.K.
Quantity
24 analyzers
Agency-reported status
Ongoing
Source record identifier
Z-2691-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityCatalog Number: C63520 UDI-DI code: 14987666545065 Serial Numbers: 2023070010 2023060005 2023090015 2023070009 2023080011 2023080012 2023100019 2023100018 2023100020 2024010037 2024010036 2023110022 2024010038 2023090016 2024020040 2024020041 2024020039 2024020043 2024020042 2023110023 2024020045 2024020046 2024010024 2024010025

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - U.S. Nationwide distribution in the states of AL, CA, GA, HI, IL, MI, MN, MS, NE, NY, OR, PA and SC. The countries of Lebanon, South Africa.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Aug 28, 2024
Recall initiation date
Jul 10, 2024
Center classification date
Aug 21, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records