Radius VSM Disposable NIBP Cuff, REF: 4825, 4826 — FDA record Z-2671-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Feb 20, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Radius VSM Disposable NIBP Cuff, REF: 4825, 4826
Reason
Two devices recalled: 1) Electrocardiogram (ECG) electrode assembly ECG feature may incorrectly detect extreme bradycardia, tachycardia and trigger alarms on normal heart rate patients, which may distract clinicians away from management of true positive alarms, which could delay treatment; 2) Small and medium size blood pressure cuffs may contain rough edges, may result in skin irritation/redness.
Classification
Class II
Recalling firm
Masimo Corporation
Quantity
1,028
Agency-reported status
Ongoing
Source record identifier
Z-2671-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityREF/UDI-DI/Lot: 4825/00843997006989/21G6S, 22JHV, 23H03, 23J05, 24DBV, 24NGG, 25ADG, E21D31, E21JWM, E23FVD; 4826/10843997006993/21HVX, 22ELB, 23G31, 23H04, 23HWL, 23J06, 24DBW, 24MNP, 24NGF, 25ADF, E21D32, E21F04, E21JFA, E21KHK

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution in the states of OK, CA , MD , MA , NY , UT , NC , TN , MN , TX , OH , PA , VA and the countries of Italy, Latvia, Iraq, Saudi Arabia, Germany, Spain, Turkey, France, Australia, Malta, Oman, United Arab Emirates, Martinique, Netherlands, Cyprus, Kuwait, Czech Republic, United Kingdom, Croatia, Switzerland, Belgium, Singapore, Romania, Colombia, Hong Kong, Israel, Finland, Poland, New Zealand, Tunisia, Chile, Algeria, Canada, Denmark, Palestinian Territory, Qatar, Greece, Paraguay.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jul 8, 2026
Recall initiation date
Feb 20, 2026
Center classification date
Jul 2, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.