Ingenuity CT CT5000 Ingenuity Plus CT5000 Ingenuity Pro CT5000 Ingenuity Premium Ingenuity CT Upgrades Ingenuity CT C... — FDA record Z-2652-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jun 11, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Ingenuity CT CT5000 Ingenuity Plus CT5000 Ingenuity Pro CT5000 Ingenuity Premium Ingenuity CT Upgrades Ingenuity CT China Ingenuity Core 128 Elite Brazil
Reason
The clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result in the clinical acquisition being inadequate for diagnosis, causing a potential for patient rescan and receiving additional radiation.
Classification
Class II
Recalling firm
PHILIPS MEDICAL SYSTEMS
Quantity
182
Agency-reported status
Ongoing
Source record identifier
Z-2652-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityFCO72800815, FCO72800816, 728321 DI (01)00884838059498 728323 DI (01)00884838059504 728326 DI (01)00884838059511 728327 DI (01)00884838098701 728324 DI (01)00884838059863 728325

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide Distribution including Puerto Rico

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Aug 5, 2026
Recall initiation date
Jun 11, 2026
Center classification date
Jul 27, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.