Recall archive / FDA device enforcement
3M Attest Super Rapid 5 Steam-Plus Steam Challenge Pack, REF 41482VF; designed to qualify and monitor dynamic-air rem... — FDA record Z-2640-2023
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Aug 22, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- 3M Attest Super Rapid 5 Steam-Plus Steam Challenge Pack, REF 41482VF; designed to qualify and monitor dynamic-air removal steam sterilization processes
- Reason
- Specific lots of 3M Attest Super Rapid Readout Biological Indicators (1492V), 3M Attest Super Rapid 5 Steam-Plus Challenge (41482V, 41482VF), and 3M Attest Super Rapid Readout Steam Challenge (1496V) may contain biological indicators with a cap that melts, deforms and/or cracks following a steam sterilization cycle of 132¿C and 135¿C.
- Classification
- Class II
- Recalling firm
- 3M Company - Health Care Business
- Quantity
- 77,280 devices
- Agency-reported status
- Ongoing
- Source record identifier
- Z-2640-2023
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | UDI/DI: 50707387787283( (shipper); Lot Codes: 41482VF, exp. 33P4ND; 33P4YN, exp. 01/02/2025 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Canada, China, Colombia, Hong Kong, Italy, Japan, Peru, Poland, South Korea, Spain, Taiwan.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Oct 4, 2023
- Recall initiation date
- Aug 22, 2023
- Center classification date
- Sep 26, 2023