Omnipod 5 Automated Insulin Delivery System — FDA record Z-2618-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Feb 28, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Omnipod 5 Automated Insulin Delivery System
Reason
An error message was received when using the Omnipod 5 App on compatible smartphones that prevented phone control users from controlling the Omnipod 5 Automated Insulin Delivery System from their Omnipod 5 App which may cause a delay in therapy. To reduce the volume of inquiries Insulet was receiving from customers, an email was sent to all phone control users and the cause of the error message was resolved by reverting to prior certificates on the Insulet cloud to secure communications between the Omnipod 5 App sotware and the Insulet Cloud. This issue was limited to only certain compatible Android smartphones and did not affect the Omnipod 5 Pods or the dedicated locked-down Controller provided by Insulet. All other Omnipod 5 users were still able to manage their insulin with the device without use of the smartphone app and affected users were able to switch to the locked-down control provided to all users when they initiated the device during the time the App was unavailable to them. Further, the Omnipod itself continued to deliver insulin as per its pre-programmed settings so although affected users who did not have the locked-down controller readily available may not have been able to command insulin boluses during this time, they continued to receive basal insulin without interruption. Lastly, users of the Omnipod 5 were advised during training to always have backup supplies ready in the event of a device malfunction which would include alternative means to deliver insulin subcutaneously until the issue is resolved to minimize disruption to their insulin regimen. There were 2,168 complaints received out of 7,838 users with the Android compatible smartphones affected and 9 MDRs were submitted, however, there were no serious injuries reported or reports of erroneous results as the primary issue was a delay in therapy. This defect cannot cause false results that could negatively impact patients.
Classification
Class II
Recalling firm
Insulet Corporation
Quantity
11,016 users
Agency-reported status
Ongoing
Source record identifier
Z-2618-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityPT-000559 Omnipod 5 App; all versions of software installed UDI-DI: 10385081120302

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Distribution throughout the US

No US distribution state could be established from this record. Check the official notice.

Dates in the source

Report date
Sep 27, 2023
Recall initiation date
Feb 28, 2023
Center classification date
Sep 20, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.