ClotTriever XL Catheter, 30 mm, REF: 41-102, used with Instructions for Use ClotTriever XL Catheter ClotTriever Throm... — FDA record Z-2587-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jul 19, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
ClotTriever XL Catheter, 30 mm, REF: 41-102, used with Instructions for Use ClotTriever XL Catheter ClotTriever Thrombectomy System, Reference: IU-01011
Reason
Thrombectomy system catheter instructions for use updated to reduce adverse events due to the following circumstances: 1) Fibrous, firmly adherent, calcified material, or suspected tumor thrombus overloading catheter bag 2) Catheter pulled through internal jugular vein access site or not pulled away from heart; may lead to device entrapment, pulmonary embolism, venous cutdown, abdominal surgery.
Classification
Class I
Recalling firm
Inari Medical - Oak Canyon
Quantity
2488
Agency-reported status
Ongoing
Source record identifier
Z-2587-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI:00850291007277. All lots. IU-01011 REV. C.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution including in the states of PA, FL, GA, CA, IL, SC, NY, MN, UT, TN, WI, MI, KS, SD, AZ, AR, IN, KY, TX, AL, MS, DE, MA, OH, MO, WV, VA, NC, NV, CO, NE, OR, NM, NH, NJ, CT, ID, IA, WA, MD, OK, RI, DC, PR, LA, ND, VT and the countries of New Zealand, Chile, Australia.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Aug 28, 2024
Recall initiation date
Jul 19, 2024
Center classification date
Aug 21, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.