Pinnacle Radiation Therapy Planning System: Multimodality Simulation Workspace 18.0.5 Model 870258 UDI code: (01)0088... — FDA record Z-2570-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Aug 5, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Pinnacle Radiation Therapy Planning System: Multimodality Simulation Workspace 18.0.5 Model 870258 UDI code: (01)00884838103566(10)18.0.5.22236 Pinnacle TumorLOC 16.2 Model 870226 UDI code: (01)00884838091122(10)16.2.0.60010 Pinnacle TumorLOC 16.2.1 Model 870227 UDI code: (01)00884838091122(10)16.2.1.60098 Pinnacle TumorLOC 16.0.2 Model 870218 UDI code: (01)00884838085602(10)16.0.2.60004 Pinnacle TumorLOC 14.0 Model 870200 UDI code: N/A Pinnacle3 Radiation Therapy Planning System is a software package intended to provide planning support for the treatment of disease processes. The full Pinnacle3 Radiation Therapy Planning System software package provides planning support for the treatment of disease processes, utilizing photon, electron and brachytherapy techniques.
Reason
Due to a software issue, there is a potential image error of the Region of Interest for expansion/contraction for HFP (Head First Prone), FFS (Feet First Supine) and FFP (Feet First Prone) orientations.
Classification
Class II
Recalling firm
Philips Medical Systems (Cleveland) Inc
Quantity
810 (350 U.S. and 460 O.U.S.)
Agency-reported status
Ongoing
Source record identifier
Z-2570-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityPinnacle¿ Radiation Therapy Planning System: Multimodality Simulation Workspace 18.0.5 Model Number: 870258 UDI code: (01)00884838103566(10)18.0.5.22236 Lot Number (Software Version): 18.0.5 Pinnacle TumorLOC 16.2 Model Number: 870226 UDI code: (01)00884838091122(10)16.2.0.60010 Lot Number (Software Version): 16.2 Pinnacle TumorLOC 16.2.1 Model Number: 870227 UDI code: (01)00884838091122(10)16.2.1.60098 Lot Number (Software Version): 16.2.1 Pinnacle TumorLOC 16.0.2 Model Number: 870218 UDI code: (01)00884838085602(10)16.0.2.60004 Lot Number (Software Version): 16.0.2 Pinnacle TumorLOC 14.0 Model Number: 870200 UDI code: N/A Lot Number (Software Version): 14.0

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

U.S Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Sep 24, 2025
Recall initiation date
Aug 5, 2025
Center classification date
Sep 16, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records