Automated Impella Controller (AIC) with the below brands and product codes. Used with heart assist pump. 1. Impella C... — FDA record Z-2568-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Aug 20, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Automated Impella Controller (AIC) with the below brands and product codes. Used with heart assist pump. 1. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 2. Impella Controller, Packaged, US, Loaner; Product Code: 0042-0000-US-L. 3. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 4. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 5. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 6. Impella Controller, UK, Loaner; Product Code: 0042-0000-UK-L.
Reason
Specific Automated Impella Controllers (AICs) that have a Pump Driver Circuit Assembly do not meet current specifications. This may lead to decreased pump performance or pump stop and trigger an Impella Failure or Impella Stopped Controller Failure alarm.
Classification
Class I
Recalling firm
Abiomed, Inc.
Quantity
71 units
Agency-reported status
Ongoing
Source record identifier
Z-2568-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicability1. Product Code: 0042-0000-US; GTIN: 813502010022; Serial Numbers: IC1026, IC1029, IC1043, IC1046, IC1040, IC1047, IC1088, IC1089, IC1093, IC1110, IC1167, IC1221, IC1225, IC1222, IC1258, IC1250, IC1272, IC1293, IC1320, IC1317, IC1353, IC1354, IC1355, IC1375, IC1377, IC1403, IC1446, IC1450, IC1452, IC1461, IC1496, IC1501, IC1508, IC1515, IC1516, IC1517, IC1518, IC1530, IC1534, IC1535, IC1539, IC1538, IC1531, IC1533, IC1540, IC1532, IC1551, IC1552, IC1550, IC1553, IC1646, IC1756, IC2255, IC2422, IC2489, IC3013. 2. Product Code: 0042-0000-US-L; GTIN: 813502010022; Serial Number: IC1281. 3. Product Code: 0042-0000-CA; GTIN: 813502011272; Serial Numbers: IC1543, IC1544. 4. Product Code: 0042-0000-EU; GTIN: 813502011289; Serial Numbers: IC1350, IC1383, IC1546, IC1548, IC1547, IC1549, IC1775, IC2243, IC4002. 5. Product Code: 0042-0000-UK; GTIN: 813502011296; Serial Numbers: IC1519, IC1542. 6. Product Code: 0042-0000-UK-L; GTIN: 813502011296; Serial Number: IC1493.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US distribution to AR, AZ, CA, CO, CT, FL, GA, ID, MA, MI, MO, MS, NJ, NY, OH, PA, TX. International distribution to Canada, Germany, Great Britian, Italy, Netherlands, Norway.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Oct 1, 2025
Recall initiation date
Aug 20, 2025
Center classification date
Sep 19, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.