Recall archive / FDA device enforcement
Reusable distal attachments Models: MH-462, MH-463, MH-464, MH-465, MH-466, MH-483, MH-587, MH-588(US), MH-589, MH-59... — FDA record Z-2523-2023
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Jul 27, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Reusable distal attachments Models: MH-462, MH-463, MH-464, MH-465, MH-466, MH-483, MH-587, MH-588(US), MH-589, MH-590, MH-591, MH-592, MH-593, MH-594, MH-595, MH-596, MH-597, MH-598, MAJ-289, MAJ-290, MAJ-291, MAJ-292, MAJ-293, MAJ-294, MAJ-295, MAJ-296, MAJ-297, MAJ-663, MAJ-664, MAJ-665, MAJ-666
- Reason
- The Distal Attachment contain di-2-ethylhexyl phthalate (DEHP), which exceeds the permissible exposure level calculated according to ISO 10993-17: 2002
- Classification
- Class II
- Recalling firm
- Aomori Olympus Co., Ltd.
- Quantity
- 1,252 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-2523-2023
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | Model: UDI: MH-462 4953170047732 MH-463 4953170047756 MH-464 4953170047770 MH-465 4953170047794 MH-466 4953170047817 MH-483 N/A MH-587 4953170047855 MH-588 4953170364518 (US) MH-589 4953170047893 MH-590 N/A MH-591 4953170047930 MH-592 N/A MH-593 4953170047978 MH-594 4953170047992 MH-595 4953170048012 MH-596 N/A MH-597 4953170048050 MH-598 N/A MAJ-289 N/A MAJ-290 N/A MAJ-291 N/A MAJ-292 N/A MAJ-293 N/A MAJ-294 N/A MAJ-295 N/A MAJ-296 N/A MAJ-297 N/A MAJ-663 N/A MAJ-664 N/A MAJ-665 N/A MAJ-666 N/A DIstal Attachment- All lots |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US Nationwide distribution.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Sep 13, 2023
- Recall initiation date
- Jul 27, 2023
- Center classification date
- Sep 7, 2023
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.